beautypg.com

B&K Precision 9130 - Manual User Manual

Page 33

background image

33

VX. XX software version number

For example: BK PRECISION, 9130, 000000000000111101,V1.66


*OPC

When all commands sent to the instrument prior to this command have been executed, OPC of the

standard event status register will be set to 1.

Command syntax: *OPC

Query syntax: *OPC ?

Example: *OPC

*PSC

This command controls if the power supply sends a query or not when it is reset.

1|ON:

When power supply is reset, operation event enable register, quest event enable register and

standard event status register are all

reset.

0|OFF:

The data of operation event enable register, Quest event enable register and standard event

status enable register is stored in a nonvolatile register, and is recalled when power supply is reset.

Command syntax: *PSC

Parameter: 0|1|ON|OFF

Return parameter: data stored

Example: *PSC ON

*RST

This command resets the power supply to its default setting.

Command syntax: *RST

Example: *RST

*SRE

This command can set the parameter of the status byte enable register. The value of this parameter determines

which bit value of the status byte register is 1 and the byte will enable RQS of status byte register is 1. The bit

definition of the status byte enable register is the same as the status byte register.

Command syntax: *SRE

Parameter: 0~255

Reset value: consult command *PSC

Example: *SRE 110

*STB?

This command reads data from the status byte register. After executing this command, the status byte register

is reset.

Command syntax: *STB?

Return parameter:

Example: *STB?

*SAV

This command saves the operating parameters of the power supply to non-volatile memory. The