7 configure voltage slope, Configure voltage slope, 0w starting profile – B&K Precision 9116 - Manual User Manual

Page 63

background image

50

10.

Once both parameters are set, press

to confirm the changed settings and press

several times to exit the menu. If the Starting profile function was setup properly,

the Trig indicator will be highlighted and Starting profile will be shown on the front

panel, shown below.

11.

To enable the Short voltage drop function, press the

key, on the front panel. The

proper voltage (12V or 24V) should be metered on the front panel display.

12.

To run the Starting profile function press

and

to trigger the test to run.

13.

Repeat step 8 to trigger the function to run as many times as needed or setup the

repeat functionality.

3.7

Configure Voltage Slope

The power supply has the capability of controlling the slope of the output voltage during

changes between voltage settings. Timings can be configured for both the rising and falling

edge between voltage output transitions.

To illustrate this feature and provide configuration instructions, follow the example below.

Suppose you want to simulate the voltage signal below by manually setting the voltage from 5 V

to 10 V over a span of 5 seconds, then manually changing it to 1 V and have it decrease from 10

V to 1 V over a span of 10 seconds. This can be done by configuring the rising and falling edge

timing.

Note: If any automotive testing is running the

will blink to indicate so. To stop

any test from running further, press

and

.

0.000V 0.000A

0.0W

Starting profile

Trig