beautypg.com

Rs-232, Usbtmc, Rs232 – B&K Precision 9116 - Manual User Manual

Page 35

background image

22

RS-232

Follow the steps below to configure the power supply for RS-232 operation:

1.

From the SYSTEM menu, browse and select Communication and press

.

2.

Select RS-232(Def) and press

to set to RS-232 for remote communication. The

following display will be shown:

3.

4800 is the baudrate; 8 is the data bits; N is the parity; 1 is the stop bit; Addr… is for

address.

4.

Use to select between each serial settings, and use to change the settings.

5.

Below lists the options that can be changed for each setting:

Baudrate: 4800, 9600, 19200, 38400, 57600, 115200*

Data bits: 8

Parity: N (None), E (Even), O (Odd)

Stop bit: 1, 2

Addr: 0 to 31

Note: The default is 4800, 8, N, 1, Address = 0.

*Setting the baudrate to 115200 may cause unstable results during remote

communication. Select a lower baudrate if communication errors occur.

6.

All serial settings must match with the settings configured on the PC in order for

communication to link successfully.

USBTMC

A USB Type A to Type B cable (i.e. USB printer cable) is required to connect the USB port in the

rear panel to a PC. Follow the steps below to setup the power supply for remote

communication.

1.

From the SYSTEM menu, browse and select Communication and press

.

2.

Select USB and press

to set to USBTMC for remote communication.

3.

Install the USB driver. For Windows® 7 users, this may install automatically. For other

users, visit

www.bkprecision.com

to download the driver.

Note: Users who have LabVIEW™ or NI-VISA installed will automatically have this

driver in their system. In this case, driver download is not required.

RS232

4800 , 8 , N , 1 , Addr . . .