beautypg.com

Profile for reset, Profile for the reset test, 0w short voltage drop – B&K Precision 9116 - Manual User Manual

Page 61

background image

48

7.

To enable the Short voltage drop function, press the

key, on the front panel. The

proper voltage (12V or 24V) should be metered on the front panel display.

8.

To run Short voltage drop function press

and

to trigger the test to run.

9.

Repeat step 8 to trigger the function to run as many times as needed.

Profile for reset

Follow the steps below to set up the ISO16750-2 Profile for the reset test function:

1.

Access the main menu by pressing

and the

.

2.

Use the key to select FUNC. and press

.

3.

Use the key to select ISO16750-2 and press

. The screen below will display:

4.

While Short voltage drop is flashing, use keys to toggle to Profile for the reset

test (shown below) and press

to access that option.

5.

Select Usmin… and press enter. The user can choose a Usmin value between 1 V – 80 V.

Once a value is selected, press

to confirm.

6.

Use the key to select Off/On and use keys to toggle between the two options.

Choose On to be able to run the function.

7.

Once both parameters are set, press

to confirm the changed settings and press

several times to exit the menu. If the Starting profile function was setup properly,

the Trig indicator will be highlighted and Profile for the reset test will be shown on the

front panel, shown below.

ISO16750-2 profile

Profile for the reset test

0.000V 0.000A

0.0W

Short voltage drop

Trig