Atec Rohde-Schwarz-ZNB Series User Manual
Page 8
Version 07.00, August 2013
8
Rohde & Schwarz R&S
®
ZNB Vector Network Analyzer
Typical sweep times versus number of measurement points
®
ZNB20 and the R&S
®
ZNB40
Number of measurement points
51
201
401
1601
5001
9 GHz start frequency, 10 GHz stop frequency, AGC LOW DIST, 1 kHz measurement bandwidth
With correction switched off
50 ms
200 ms
370 ms
1450 ms
4500 ms
With 2-port TOSM calibration
100 ms
400 ms
740 ms
2900 ms
9000 ms
With 4-port TOSM calibration
200 ms
800 ms
1480 ms
5800 ms
18000 ms
9 GHz start frequency, 10 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth
With correction switched off
2 ms
4.5 ms
8 ms
25 ms
75 ms
With 2-port TOSM calibration
4 ms
9 ms
16 ms
50 ms
150 ms
With 4-port TOSM calibration
8 ms
18 ms
32 ms
100 ms
300 ms
9 GHz start frequency, 10 GHz stop frequency, AGC AUTO, 1 MHz measurement bandwidth
With correction switched off
1.5 ms
3.2 ms
5 ms
10 ms
25 ms
With 2-port TOSM calibration
3 ms
6.5 ms
10 ms
22 ms
63 ms
With 4-port TOSM calibration
7 ms
13 ms
20 ms
56 ms
165 ms
10 GHz start frequency, 20 GHz stop frequency, AGC LOW DIST, 1 kHz measurement bandwidth
With correction switched off
51 ms
190 ms
370 ms
1500 ms
4600 ms
With 2-port TOSM calibration
102 ms
380 ms
740 ms
3000 ms
9100 ms
With 4-port TOSM calibration
205 ms
760 ms
1470 ms
6000 ms
18500 ms
10 GHz start frequency, 20 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth
With correction switched off
6 ms
10 ms
15 ms
32 ms
100 ms
With 2-port TOSM calibration
12 ms
20 ms
30 ms
65 ms
200 ms
With 4-port TOSM calibration
24 ms
40 ms
60 ms
130 ms
400 ms
10 GHz start frequency, 20 GHz stop frequency, AGC AUTO, 1 MHz measurement bandwidth
With correction switched off
5 ms
8 ms
10 ms
20 ms
45 ms
With 2-port TOSM calibration
10 ms
16 ms
20 ms
40 ms
90 ms
With 4-port TOSM calibration
21 ms
32 ms
40 ms
85 ms
220 ms
100 kHz start frequency, 20 GHz stop frequency, AGC LOW DIST, 1 kHz measurement bandwidth
With correction switched off
55 ms
200 ms
380 ms
1500 ms
4600 ms
With 2-port TOSM calibration
115 ms
400 ms
760 ms
3000 ms
9100 ms
With 4-port TOSM calibration
230 ms
800 ms
1520 ms
6000 ms
18500 ms
100 kHz start frequency, 20 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth
With correction switched off
10 ms
15 ms
19 ms
40 ms
100 ms
With 2-port TOSM calibration
20 ms
30 ms
38 ms
80 ms
200 ms
With 4-port TOSM calibration
40 ms
60 ms
75 ms
160 ms
400 ms
100 kHz start frequency, 20 GHz stop frequency, AGC AUTO, 1 MHz measurement bandwidth
With correction switched off
9 ms
14 ms
16 ms
30 ms
60 ms
With 2-port TOSM calibration
18 ms
28 ms
32 ms
60 ms
120 ms
With 4-port TOSM calibration
38 ms
56 ms
64 ms
120 ms
250 ms
10 MHz start frequency, 40 GHz stop frequency, AGC LOW DIST, 1 kHz measurement bandwidth
With correction switched off
60 ms
200 ms
380 ms
1500 ms
4600 ms
With 2-port TOSM calibration
120 ms
400 ms
760 ms
3000 ms
9100 ms
10 MHz start frequency, 40 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth
With correction switched off
12 ms
20 ms
25 ms
45 ms
100 ms
With 2-port TOSM calibration
24 ms
40 ms
50 ms
90 ms
200 ms
10 MHz start frequency, 40 GHz stop frequency, AGC AUTO, 1 MHz measurement bandwidth
With correction switched off
11 ms
18 ms
20 ms
35 ms
65 ms
With 2-port TOSM calibration
22 ms
36 ms
40 ms
70 ms
130 ms
4
Sweep time is to be understood as cycle time; static frequency accuracy of the instrument applies; measured with firmware version 1.90, Windows 7.