R&s®znb-b81, Znb-b81 – Atec Rohde-Schwarz-ZNB Series User Manual

Page 24

background image

Version 07.00, August 2013

24

Rohde & Schwarz R&S

®

ZNB Vector Network Analyzer

Trace stability

Trace noise magnitude (RMS)

at 0 dBm source power,

0 dB reflection

IF bandwidth

9 kHz to 20 kHz

1 kHz

< 0.008 dB, typ. 0.004 dB

20 kHz to 100 kHz

1 kHz

< 0.005 dB, typ. 0.001 dB

100 kHz to 100 MHz

10 kHz

< 0.005 dB, typ. 0.001 dB

100 MHz to 8.5 GHz

10 kHz

< 0.005 dB, typ. 0.002 dB

Measurement speed

Typical sweep times versus number of measurement points

Number of measurement points

51

201

401

1601

5001

800 MHz start frequency, 1 GHz stop frequency, AGC LOW DIST, 100 kHz measurement bandwidth

With correction switched off

2.0 ms

5 ms

8 ms

20 ms

57 ms

With 2-port TOSM calibration

3.5 ms

9 ms

13 ms

40 ms

113 ms

With 4-port TOSM calibration

6.5 ms

17 ms

25 ms

81 ms

246 ms

800 MHz start frequency, 1 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth

With correction switched off

3.5 ms

10 ms

16 ms

55 ms

170 ms

With 2-port TOSM calibration

6 ms

18 ms

31 ms

109 ms

339 ms

With 4-port TOSM calibration

10 ms

35 ms

61 ms

225 ms

701 ms

100 kHz start frequency, 4.5 GHz stop frequency, AGC LOW DIST, 100 kHz measurement bandwidth

With correction switched off

4.0 ms

8 ms

12 ms

33 ms

90 ms

With 2-port TOSM calibration

7.5 ms

14 ms

22 ms

65 ms

180 ms

With 4-port TOSM calibration

14 ms

27 ms

42 ms

130 ms

355 ms

100 kHz start frequency, 4.5 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth

With correction switched off

6 ms

12 ms

21 ms

69 ms

205 ms

With 2-port TOSM calibration

10 ms

23 ms

40 ms

137 ms

405 ms

With 4-port TOSM calibration

19 ms

45 ms

79 ms

273 ms

810 ms

100 kHz start frequency, 8.5 GHz stop frequency, AGC LOW DIST, 100 kHz measurement bandwidth

With correction switched off

4.5 ms

9 ms

13 ms

34 ms

90 ms

With 2-port TOSM calibration

8.5 ms

17 ms

25 ms

67 ms

180 ms

With 4-port TOSM calibration

16 ms

32 ms

47 ms

131 ms

359 ms

100 kHz start frequency, 8.5 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth

With correction switched off

6 ms

13 ms

22 ms

70 ms

205 ms

With 2-port TOSM calibration

11 ms

26 ms

43 ms

139 ms

410 ms

With 4-port TOSM calibration

21 ms

50 ms

84 ms

280 ms

815 ms

The R&S

®

ZNBx-B52/-B54 options cannot be combined with the R&S

®

ZNBx-B1 option and/or

the R&S

®

ZNBx-B31/-B32/-B33/-B34 options.

R&S

®

ZNB-B81

This data is valid from +18 °C to +28 °C and a measurement bandwidth at maximum 10 kHz.

DC inputs

Number of ports

4

Connector type

BNC, female

Voltage range

±20 V, ±3 V, ±0.3 V

Measurement accuracy

±20 V

1 % of reading + 0.01 V

±3 V

1 % of reading + 0.001 V

±0.3 V

1 % of reading ± 0.001 V

Input impedance

≥ 1 MΩ

Damage voltage

30 V