beautypg.com

Measurement accuracy of the r&s®znb40, Measurement accuracy of the r&s, Znb40 – Atec Rohde-Schwarz-ZNB Series User Manual

Page 11: Znb vector network analyzer 11

background image

Version 07.00, August 2013

Rohde & Schwarz R&S

®

ZNB Vector Network Analyzer

11

Measurement accuracy of the R&S

®

ZNB40

This data is valid between +18 °C and +28 °C, provided the temperature has not varied by more than 1 °C after calibration. Validity of

the data is conditional on the use of an R&S

®

ZV-Z229 calibration kit. This calibration kit is used to achieve the effective system data

specified below. Frequency points, measurement bandwidth and sweep time have to be identical for measurement and calibration

(no interpolation allowed).

Accuracy of transmission measurements

10 MHz to 200 MHz

0 dB to –35 dB

< 0.07 dB or < 1°

–35 dB to –50 dB

< 0.1 dB or < 1°

–50 dB to –60 dB

< 0.2 dB or < 2°

200 MHz to 20 GHz

0 dB to –35 dB

< 0.07 dB or < 1°

–35 dB to –50 dB

< 0.1 dB or < 1°

–50 dB to –60 dB

< 0.1 dB or < 2°

20 GHz to 40 GHz

0 dB to –35 dB

< 0.1 dB or < 1°

–35 dB to –50 dB

< 0.1 dB or < 1°

–50 dB to –60 dB

< 0.2 dB or < 2°

Specifications are based on a matched DUT, a measurement bandwidth of 10 Hz and a nominal source power of –10 dBm.

0,01

0,10

1,00

10,00

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient |S21| in dB

M

ag

ni

tu

de

U

nc

er

ta

in

ty

in

d

B

10MHz to 200MHz
200MHz to 20GHz
20GHz to 40GHz

0,1

1,0

10,0

100,0

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient |S21| in dB

Ph

ase u

ncer

tai

nt

y i

n d

eg

rees

10MHz to 200MHz
200MHz to 20GHz
20GHz to 40GHz

Typical accuracy of transmission magnitude and transmission phase measurements for the R&S

®

ZNB40 in the frequency range

from 10 MHz to 40 GHz. Analysis conditions: S

11

= S

22

= 0, cal. power –10 dBm, meas. power –10 dBm.

Accuracy of reflection measurements

10 MHz to 4 GHz

0 dB to –15 dB

< 0.5 dB or < 2.5°

–15 dB to –25 dB

< 1.0 dB or < 6.5°

–25 dB to –35 dB

< 3.8 dB or < 20°

4 GHz to 20 GHz

0 dB to –15 dB

< 0.7 dB or < 4°

–15 dB to –25 dB

< 1.8 dB or < 11°

–25 dB to –35 dB

< 7.5 dB or < 35°

20 GHz to 40 GHz

0 dB to –15 dB

< 1.0 dB or < 6°

–15 dB to –25 dB

< 3.0 dB or < 17°

–25 dB to –35 dB

< 17 dB or < 60°

Specifications are based on an isolating DUT, a measurement bandwidth of 10 Hz and a nominal source power of –10 dBm.

0,00

0,01

0,02

0,03

0,04

0,05

0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1

Reflection coefficient |S11| (linear)

M

ag

ni

tu

de u

ncer

tai

nt

y (

lin

ear

)

10MHz to 4GHz
4GHz to 20GHz
20GHz to 40GHz

0,0

1,0

2,0

3,0

4,0

5,0

6,0

7,0

8,0

9,0

10,0

0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1

Reflection coefficient |S11| (linear)

Ph

ase u

ncer

tai

nt

y (

lin

ear

)

10MHz to 4GHz
4GHz to 20GHz
20GHz to 40GHz

Typical accuracy of reflection magnitude and reflection phase measurements for the R&S

®

ZNB40 in the frequency range

from 10 MHz to 40 GHz. Analysis conditions: S

12

= S

21

= 0, cal. power –10 dBm, meas. power –10 dBm.