beautypg.com

Measurement accuracy of the r&s®znb20, Measurement accuracy of the r&s, Znb20 – Atec Rohde-Schwarz-ZNB Series User Manual

Page 10: Znb vector network analyzer

background image

Version 07.00, August 2013

10

Rohde & Schwarz R&S

®

ZNB Vector Network Analyzer

Measurement accuracy of the R&S

®

ZNB20

This data is valid between +18 °C and +28 °C, provided the temperature has not varied by more than 1 °C after calibration. Validity of

the data is conditional on the use of an R&S

®

ZV-Z235 calibration kit. This calibration kit is used to achieve the effective system data

specified below. Frequency points, measurement bandwidth and sweep time have to be identical for measurement and calibration

(no interpolation allowed).

Accuracy of transmission measurements

100 kHz to 1 MHz

+5 dB to –35 dB

< 0.05 dB or < 0.5°

–35 dB to –50 dB

< 0.1 dB or < 1°

–50 dB to –60 dB

< 0.3 dB or < 5°

1 MHz to 10 GHz

+5 dB to –35 dB

< 0.05 dB or < 0.5°

–35 dB to –50 dB

< 0.06 dB or < 0.6°

–50 dB to –60 dB

< 0.1 dB or < 1°

10 GHz to 20 GHz

+5 dB to –35 dB

< 0.05 dB or < 0.5°

–35 dB to –50 dB

< 0.08 dB or < 0.8°

–50 dB to –60 dB

< 0.15 dB or < 1.5°

Specifications are based on a matched DUT, a measurement bandwidth of 10 Hz and a nominal source power of –10 dBm.

0,01

0,10

1,00

10,00

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient |S21| in dB

M

ag

ni

tu

de

U

nc

er

ta

in

ty

in

d

B

100kHz to 1MHz
1MHz to 10GHz
10GHz to 20GHz

0,1

1,0

10,0

100,0

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient |S21| in dB

Ph

ase u

ncer

tai

nt

y i

n d

eg

rees

100kHz to 1MHz
1MHz to 10GHz
10GHz to 20GHz

Typical accuracy of transmission magnitude and transmission phase measurements for the R&S

®

ZNB20 in the frequency range

from 100 kHz to 20 GHz. Analysis conditions: S

11

= S

22

= 0, cal. power –10 dBm, meas. power –10 dBm.

Accuracy of reflection measurements

100 kHz to 10 GHz

0 dB to –15 dB

< 0.3 dB or < 2°

–15 dB to –25 dB

< 1.0 dB or < 6°

–25 dB to –35 dB

< 3.0 dB or < 20°

10 GHz to 20 GHz

0 dB to –15 dB

< 0.5 dB or < 4°

–15 dB to –25 dB

< 1.6 dB or < 10°

–25 dB to –35 dB

< 7.5 dB or < 35°

Specifications are based on an isolating DUT, a measurement bandwidth of 10 Hz and a nominal source power of –10 dBm.

0,00

0,01

0,02

0,03

0,04

0,05

0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1

Reflection coefficient |S11| (linear)

M

ag

ni

tu

de u

ncer

tai

nt

y (

lin

ear

)

100kHz to 10GHz

10GHz to 20GHz

0,0

1,0

2,0

3,0

4,0

5,0

6,0

7,0

8,0

9,0

10,0

0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1

Reflection coefficient |S11| (linear)

Ph

ase u

ncer

tai

nt

y (

lin

ear

)

100kHz to 10GHz

10GHz to 20GHz

Typical accuracy of reflection magnitude and reflection phase measurements for the R&S

®

ZNB20 in the frequency range

from 100 kHz to 20 GHz. Analysis conditions: S

12

= S

21

= 0, cal. power –10 dBm, meas. power –10 dBm.