Version 07.00, August 2013

Rohde & Schwarz R&S

®

ZNB Vector Network Analyzer

3

Definitions

General

Product data applies under the following conditions:

Three hours storage at ambient temperature followed by 60 minutes warm-up operation

Specified environmental conditions met

Recommended calibration interval adhered to

All internal automatic adjustments performed, if applicable

Unless stated otherwise, specifications apply to test ports and a nominal source power of –10 dBm

Specifications with limits

Represent warranted product performance by means of a range of values for the specified parameter. These specifications are

marked with limiting symbols such as <, ≤, >, ≥, ±, or descriptions such as maximum, limit of, minimum. Compliance is ensured by

testing or is derived from the design. Test limits are narrowed by guard bands to take into account measurement uncertainties, drift

and aging, if applicable.

Specifications without limits

Represent warranted product performance for the specified parameter. These specifications are not specially marked and represent

values with no or negligible deviations from the given value (e.g. dimensions or resolution of a setting parameter). Compliance is

ensured by design.

Typical data (typ.)

Characterizes product performance by means of representative information for the given parameter. When marked with <, > or as a

range, it represents the performance met by approximately 80 % of the instruments at production time. Otherwise, it represents the

mean value.

Nominal values (nom.)

Characterize product performance by means of a representative value for the given parameter (e.g. nominal impedance). In contrast to

typical data, a statistical evaluation does not take place and the parameter is not tested during production.

Measured values (meas.)

Characterize expected product performance by means of measurement results gained from individual samples.

Uncertainties

Represent limits of measurement uncertainty for a given measurand. Uncertainty is defined with a coverage factor of 2 and has been

calculated in line with the rules of the Guide to the Expression of Uncertainty in Measurement (GUM), taking into account

environmental conditions, aging, wear and tear.

Device settings and GUI parameters are indicated as follows: “parameter: value”.

Typical data as well as nominal and measured values are not warranted by Rohde & Schwarz.