beautypg.com

Test port input – Atec Rohde-Schwarz-ZNB Series User Manual

Page 19

background image

Version 07.00, August 2013

Rohde & Schwarz R&S

®

ZNB Vector Network Analyzer

19

Test port input

Match

without system error correction

R&S

®

ZNB4

9 kHz to 20 kHz

> 10 dB

20 kHz to 4.5 GHz

> 20 dB

R&S

®

ZNB8

9 kHz to 20 kHz

> 10 dB

20 kHz to 8.5 GHz

> 20 dB

R&S

®

ZNB20

100 kHz to 1 MHz

> 16 dB

1 MHz to 2 GHz

> 20 dB

2 GHz to 20 GHz

> 16 dB

R&S

®

ZNB40

10 MHz to 50 MHz

> 15 dB

50 MHz to 2 GHz

> 20 dB

2 GHz to 6 GHz

> 16 dB

6 GHz to 10 GHz

> 12 dB

10 GHz to 20 GHz

> 10 dB

20 GHz to 40 GHz

> 8 dB

Maximum nominal input level

+13 dBm

Power measurement accuracy

at –10 dBm without power calibration

R&S

®

ZNB4 and R&S

®

ZNB8

9 kHz to 100 kHz

< 2 dB

100 kHz to 8.5 GHz

< 1 dB

R&S

®

ZNB20

100 kHz to 20 GHz

< 1 dB

R&S

®

ZNB40

10 MHz to 20 GHz

< 1 dB

20 GHz to 40 GHz

< 2 dB

Receiver linearity

referenced to –10 dBm

R&S

®

ZNB4 and R&S

®

ZNB8

for +20 dB to +10 dB

9 kHz to 7.5 GHz

< 0.2 dB

for +18 dB to +10 dB

7.5 GHz to 8.5 GHz

< 0.2 dB

for +10 dB to –40 dB

9 kHz to 8.5 GHz

< 0.1 dB

R&S

®

ZNB20

for +18 dB to +10 dB

100 kHz to 1 MHz

< 0.3 dB

for +20 dB to +10 dB

1 MHz to 15 GHz

< 0.3 dB

for +18 dB to +10 dB

15 GHz to 20 GHz

< 0.3 dB

for +10 dB to -40 dB

100 kHz to 20 GHz

< 0.1 dB

R&S

®

ZNB40

for +15 dB to +10 dB

10 MHz to 40 GHz

< 0.4 dB

for +10 dB to +5 dB

10 MHz to 40 GHz

< 0.2 dB

for +5 dB to -40 dB

10 MHz to 40 GHz

< 0.1 dB

Damage level

+27 dBm

Damage DC voltage

30 V