Test port input – Atec Rohde-Schwarz-ZNB Series User Manual
Page 19
Version 07.00, August 2013
Rohde & Schwarz R&S
®
ZNB Vector Network Analyzer
19
Test port input
Match
without system error correction
R&S
®
ZNB4
9 kHz to 20 kHz
> 10 dB
20 kHz to 4.5 GHz
> 20 dB
R&S
®
ZNB8
9 kHz to 20 kHz
> 10 dB
20 kHz to 8.5 GHz
> 20 dB
R&S
®
ZNB20
100 kHz to 1 MHz
> 16 dB
1 MHz to 2 GHz
> 20 dB
2 GHz to 20 GHz
> 16 dB
R&S
®
ZNB40
10 MHz to 50 MHz
> 15 dB
50 MHz to 2 GHz
> 20 dB
2 GHz to 6 GHz
> 16 dB
6 GHz to 10 GHz
> 12 dB
10 GHz to 20 GHz
> 10 dB
20 GHz to 40 GHz
> 8 dB
Maximum nominal input level
+13 dBm
Power measurement accuracy
at –10 dBm without power calibration
R&S
®
ZNB4 and R&S
®
ZNB8
9 kHz to 100 kHz
< 2 dB
100 kHz to 8.5 GHz
< 1 dB
R&S
®
ZNB20
100 kHz to 20 GHz
< 1 dB
R&S
®
ZNB40
10 MHz to 20 GHz
< 1 dB
20 GHz to 40 GHz
< 2 dB
Receiver linearity
referenced to –10 dBm
R&S
®
ZNB4 and R&S
®
ZNB8
for +20 dB to +10 dB
9 kHz to 7.5 GHz
< 0.2 dB
for +18 dB to +10 dB
7.5 GHz to 8.5 GHz
< 0.2 dB
for +10 dB to –40 dB
9 kHz to 8.5 GHz
< 0.1 dB
R&S
®
ZNB20
for +18 dB to +10 dB
100 kHz to 1 MHz
< 0.3 dB
for +20 dB to +10 dB
1 MHz to 15 GHz
< 0.3 dB
for +18 dB to +10 dB
15 GHz to 20 GHz
< 0.3 dB
for +10 dB to -40 dB
100 kHz to 20 GHz
< 0.1 dB
R&S
®
ZNB40
for +15 dB to +10 dB
10 MHz to 40 GHz
< 0.4 dB
for +10 dB to +5 dB
10 MHz to 40 GHz
< 0.2 dB
for +5 dB to -40 dB
10 MHz to 40 GHz
< 0.1 dB
Damage level
+27 dBm
Damage DC voltage
30 V