Atec Rohde-Schwarz-ZNB Series User Manual
Page 7
Version 07.00, August 2013
Rohde & Schwarz R&S
®
ZNB Vector Network Analyzer
7
Typical sweep times versus number of measurement points
®
ZNB4 and the R&S
®
ZNB8
Number of measurement points
51
201
401
1601
5001
800 MHz start frequency, 1 GHz stop frequency, AGC LOW DIST, 1 kHz measurement bandwidth
With correction switched off
49 ms
185 ms
370 ms
1385 ms
4320 ms
With 2-port TOSM calibration
98 ms
370 ms
740 ms
2770 ms
8640 ms
With 4-port TOSM calibration
190 ms
735 ms
1460 ms
5530 ms
17200 ms
800 MHz start frequency, 1 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth
With correction switched off
2.0 ms
5 ms
8 ms
20 ms
57 ms
With 2-port TOSM calibration
3.5 ms
9 ms
13 ms
40 ms
113 ms
With 4-port TOSM calibration
6.5 ms
17 ms
25 ms
81 ms
246 ms
800 MHz start frequency, 1 GHz stop frequency, AGC AUTO, 1 MHz measurement bandwidth
With correction switched off
1.5 ms
2.5 ms
3.5 ms
9 ms
23 ms
With 2-port TOSM calibration
2.5 ms
4 ms
6 ms
17 ms
48 ms
With 4-port TOSM calibration
4.6 ms
9 ms
13 ms
41 ms
117 ms
100 kHz start frequency, 4.5 GHz stop frequency, AGC LOW DIST, 1 kHz measurement bandwidth
With correction switched off
52 ms
185 ms
365 ms
1435 ms
4480 ms
With 2-port TOSM calibration
103 ms
370 ms
730 ms
2870 ms
8960 ms
With 4-port TOSM calibration
200 ms
735 ms
1450 ms
5750 ms
17800 ms
100 kHz start frequency, 4.5 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth
With correction switched off
4.0 ms
8 ms
12 ms
33 ms
90 ms
With 2-port TOSM calibration
7.5 ms
14 ms
22 ms
65 ms
180 ms
With 4-port TOSM calibration
14 ms
27 ms
42 ms
130 ms
355 ms
100 kHz start frequency, 4.5 GHz stop frequency, AGC AUTO, 1 MHz measurement bandwidth
With correction switched off
3.5 ms
6.0 ms
8 ms
21 ms
45 ms
With 2-port TOSM calibration
7.0 ms
11.0 ms
16 ms
42 ms
90 ms
With 4-port TOSM calibration
12 ms
20 ms
30 ms
84 ms
180 ms
100 kHz start frequency, 8.5 GHz stop frequency, AGC LOW DIST, 1 kHz measurement bandwidth
With correction switched off
53 ms
188 ms
368 ms
1435 ms
4450 ms
With 2-port TOSM calibration
105 ms
376 ms
736 ms
2870 ms
8900 ms
With 4-port TOSM calibration
208 ms
745 ms
1470 ms
5750 ms
17750 ms
100 kHz start frequency, 8.5 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth
With correction switched off
4.5 ms
9 ms
13 ms
34 ms
90 ms
With 2-port TOSM calibration
8.5 ms
17 ms
25 ms
67 ms
180 ms
With 4-port TOSM calibration
16 ms
32 ms
47 ms
131 ms
359 ms
100 kHz start frequency, 8.5 GHz stop frequency, AGC AUTO, 1 MHz measurement bandwidth
With correction switched off
3.9 ms
7 ms
10 ms
22 ms
55 ms
With 2-port TOSM calibration
8.0 ms
14 ms
20 ms
43 ms
110 ms
With 4-port TOSM calibration
15 ms
25 ms
35 ms
86 ms
214 ms
3
Sweep time is to be understood as cycle time; static frequency accuracy of the instrument applies; measured with firmware version 1.81, Windows 7.