R&s®znb4-b52/-b54 and r&s®znb8-b52/-b54, Ole_link2, Znb40-b22 – Atec Rohde-Schwarz-ZNB Series User Manual
Page 23

Version 07.00, August 2013
Rohde & Schwarz R&S
®
ZNB Vector Network Analyzer
23
R&S
®
ZNB4-B22/-B24, R&S
®
ZNB8-B22/-B24, R&S
®
ZNB20-B22/-B24 and
R&S
®
ZNB40-B22
Extended power range
Frequency range
R&S
®
ZNB4-B22 and R&S
®
ZNB4-B24
9 kHz to 4.5 GHz
R&S
®
ZNB8-B22 and R&S
®
ZNB8-B24
9 kHz to 8.5 GHz
R&S
®
ZNB20-B22 and R&S
®
ZNB20-B24
100 kHz to 20 GHz
R&S
®
ZNB40-B22
10 MHz to 40 GHz
Power range for the R&S
®
ZNB4 and
the R&S
®
ZNB8
9 kHz to 100 MHz
–85 dBm to +10 dBm, typ. +12 dBm
100 MHz to 2.5 GHz
–85 dBm to +13 dBm, typ. +15 dBm
2.5 GHz to 7.5 GHz
–85 dBm to +10 dBm, typ. +13 dBm
7.5 GHz to 8.5 GHz
–85 dBm to +8 dBm, typ. +12 dBm
Power range for the R&S
®
ZNB20
100 kHz to 1 MHz
–60 dBm to +8 dBm, typ. +10 dBm
1 MHz to 10 MHz
–60 dBm to +10 dBm, typ. +12 dBm
10 MHz to 10 GHz
–60 dBm to +12 dBm, typ. +14 dBm
10 GHz to 15 GHz
–60 dBm to +10 dBm, typ. +12 dBm
15 GHz to 20 GHz
–60 dBm to +8 dBm, typ. +10 dBm
Power range for the R&S
®
ZNB40
10 MHz to 30 GHz
–60 dBm to +10 dBm, typ. +15 dBm
30 GHz to 40 GHz
–60 dBm to +8 dBm, typ. +13 dBm
R&S
®
ZNB4-B31/-B32/-B33/-B34 and R&S
®
ZNB8-B31/-B32/-B33/-B34
Receiver step attenuators
Frequency range
R&S
®
ZNB4-B31/-B32/-B33/-B34
9 kHz to 4.5 GHz
R&S
®
ZNB8-B31/-B32/-B33/-B34
9 kHz to 8.5 GHz
Attenuation
0 dB to 30 dB in 10 dB steps
R&S
®
ZNB4-B52/-B54 and R&S
®
ZNB8-B52/-B54
Extended dynamic range
Power range
without R&S
®
ZNB4-B22/-B24 or R&S
®
ZNB8-B22/-B24 extended power range option
9 kHz to 100 kHz
–55 dBm to +8 dBm
100 kHz to 6.5 GHz
–55 dBm to +10 dBm
6.5 GHz to 7.5 GHz
–55 dBm to +8 dBm
7.5 GHz to 8.5 GHz
–55 dBm to +6 dBm
with R&S
®
ZNB4-B22/-B24 or R&S
®
ZNB8-B22/-B24 extended power range option
9 kHz to 100 kHz
–85 dBm to +8 dBm
100 kHz to 6.5 GHz
–85 dBm to +10 dBm
6.5 GHz to 7.5 GHz
–85 dBm to +8 dBm
7.5 GHz to 8.5 GHz
–85 dBm to +6 dBm
Dynamic range
9 kHz to 100 kHz
> 120 dB, typ. 130 dB
100 kHz to 50 MHz
> 125 dB, typ. 140 dB
50 MHz to 6.5 GHz
> 140 dB, typ. 150 dB
6.5 GHz to 8.5 GHz
> 130 dB, typ. 138 dB
Test port input
Match
without system error correction
9 kHz to 20 kHz
> 10 dB
20 kHz to 8.5 GHz
> 18 dB
Maximum nominal input level
+10 dBm
Receiver linearity
referenced to –10 dBm
for 18 dB to +10 dB
9 kHz to 7.5 GHz
< 0.2 dB
for +16 dB to +10 dB
7.5 GHz to 8.5 GHz
< 0.2 dB
for +10 dB to –40 dB
9 kHz to 8.5 GHz
< 0.1 dB
Noise level
at 1 kHz measurement bandwidth,
normalized to 1 Hz
9 kHz to 50 kHz
< –125 dBm (1 Hz)
50 kHz to 50 MHz
< –130 dBm (1 Hz)
50 MHz to 6.5 GHz
< –140 dBm (1 Hz)
6.5 GHz to 8.5 GHz
< –130 dBm (1 Hz)
The noise level is defined as the RMS value of the specified noise floor.