beautypg.com

Infiniimax i series performance characteristics – Atec Agilent-DSA90404A User Manual

Page 31

background image

31

InfiniiMax I Series

Performance characteristics

1134A, 1132A, 1131A, 1130A

Bandwidth*

1134A: > 7 GHz

1131A: > 3.5 GHz

1132A: > 5 GHz

1130A: > 1.5 GHz

Rise and fall time (10% to 90%)

1134A: 60 ps

1131A: 100 ps

1132A: 86 ps

1130A: 233 ps

System bandwidth (–3 dB)

1134A w/90604A: 6 GHz

1132A w/90404A: 4 GHz

1131A w/90254A: 2.5 GHz

Input capacitance

1

Cm = 0.10 pF Cm is between tips

Cg = 0.34 pF

Cg is to ground for each tip

Cdiff = 0.27 pF Differential mode capacitance = Cm + Cg/2

Cse = 0.44 pF Single-ended mode capacitance = Cm + Cg

Input resistance*

Differential mode resistance = 50 kΩ ± 2%

Single-ended mode resistance = 25 kΩ ± 2%

Input dynamic range

5.0 V peak to peak, ± 2.5 V

Input common mode range

6.75 V peak to peak dc to 100 Hz; 1.25 V peak to peak > 100 Hz

Maximum signal slew rate

18 V/ns when probing a single-ended signal

30 V/ns when probing a differential signal

DC attenuation

10:1 ± 3% before calibration on oscilloscope

10:1 ± 1% after calibration on oscilloscope

Zero offset error referred to input

< 30 mV before calibration on oscilloscope

< 5 mV after calibration on oscilloscope

Offset range

± 12.0 V when probing single-ended

Offset accuracy

< ± 1% of setting when probing single-ended

Noise referred to input

3.0 mV rms

Propagation delay

~6 ns (this delay can be deskewed relative to other signals)

Maximum input voltage

30 V peak, CAT I

ESD tolerance

> 8 kV from 100 pF, 300 Ω HBM

Temperature

Operating: 5 °C to +40 °C

Non-operating: 0 °C to +70 °C

* Denotes warranted specifications, all others are typical.

1 Measured using the probe amplifier and solder-in differential probe head with full bandwidth resistors.