InfiniiMax I Series
Performance characteristics
1134A, 1132A, 1131A, 1130A
Bandwidth*
1134A: > 7 GHz
1131A: > 3.5 GHz
1132A: > 5 GHz
1130A: > 1.5 GHz
Rise and fall time (10% to 90%)
1134A: 60 ps
1131A: 100 ps
1132A: 86 ps
1130A: 233 ps
System bandwidth (–3 dB)
1134A w/90604A: 6 GHz
1132A w/90404A: 4 GHz
1131A w/90254A: 2.5 GHz
Input capacitance
1
Cm = 0.10 pF Cm is between tips
Cg = 0.34 pF
Cg is to ground for each tip
Cdiff = 0.27 pF Differential mode capacitance = Cm + Cg/2
Cse = 0.44 pF Single-ended mode capacitance = Cm + Cg
Input resistance*
Differential mode resistance = 50 kΩ ± 2%
Single-ended mode resistance = 25 kΩ ± 2%
Input dynamic range
5.0 V peak to peak, ± 2.5 V
Input common mode range
6.75 V peak to peak dc to 100 Hz; 1.25 V peak to peak > 100 Hz
Maximum signal slew rate
18 V/ns when probing a single-ended signal
30 V/ns when probing a differential signal
DC attenuation
10:1 ± 3% before calibration on oscilloscope
10:1 ± 1% after calibration on oscilloscope
Zero offset error referred to input
< 30 mV before calibration on oscilloscope
< 5 mV after calibration on oscilloscope
Offset range
± 12.0 V when probing single-ended
Offset accuracy
< ± 1% of setting when probing single-ended
Noise referred to input
3.0 mV rms
Propagation delay
~6 ns (this delay can be deskewed relative to other signals)
Maximum input voltage
30 V peak, CAT I
ESD tolerance
> 8 kV from 100 pF, 300 Ω HBM
Temperature
Operating: 5 °C to +40 °C
Non-operating: 0 °C to +70 °C
* Denotes warranted specifications, all others are typical.