beautypg.com

Atec Agilent-DSA90404A User Manual

Page 29

background image

29

Infiniium 90000A Series Oscilloscopes

Performance characteristics

Computer system and peripherals, I/O ports (continued)

I/O ports (continued)

PCI Express

GPIB
RS-232 (serial)
Parallel
PS/2
USB 2.0 hi-speed (host)
USB 2.0 hi-speed (device)
Dual-monitor video output
Auxiliary output
Trigger output
Time base reference output

Time base reference input
LXI compliance

PCI Express x4 link, enabled by sockets (optional- Option 823)

IEEE 488.2, fully programmable (optional – Option 805)
COM1, printer and pointing device support
Centronics printer port
Two ports. Supports PS/2 pointing and input devices.
Three USB 2.0 hi-speed host ports on front panel plus four USB 2.0 Hi-Speed host ports on rear panel
One USB 2.0 hi-speed device port on rear panel that enables USB instrument control
15 pin XGA (1024x768), full color output of scope waveform display or dual monitor video output
DC (± 2.4 V); square wave (~715 Hz and ~456 MHz); trigger output (255 mV p-p into 50)
5 V 50 Ω back-terminated
10 MHz filtered sine wave with all harmonics ≤ –40 dBc. Amplitude into 50 Ω: 800 mV p-p to

1.26 V p-p (4 dBm ± 2 dB) if derived from internal reference. Tracks external reference input

amplitude ± 1 dB if applied and selected.
Must be 10 MHz, input Z0 = 50 Ω. Minimum 500 mV p-p (–2 dBm), maximum 2.0 V p-p (+10 dBm).
Functional Class C

General characteristics

Temperature

11

Operating: 5 °C to +40 °C; Non-operating: –40 °C to +65 °C

Humidity

Operating: up to 95% relative humidity (non-condensing) at +40 °C; Non-operating: up to 90% relative

humidity at +65 °C

Altitude

Operating: up to 4,000 meters (12,000 feet); Non-operating: up to 15,300 meters (50,000 feet)

Vibration

Under vibration: for operating random the 0.3 g(rms) should be 0.21 g(rms), for non-operating random

the 2.41 g(rms) should be 2.0 g(rms) and for swept sins the (0.75g) should be (0.50g).

Power

100 - 240 VAC at 50/60 Hz; maximum input power 800 Watts

Weight

Net: 20 kg (44 lbs.)

Shipping: 27.4 kg (60 lbs.)

Dimensions (excluding handle)

Height: 283 mm (11.13 inch); Width: 432 mm (17.02 inch); Depth: 506 mm (19.91 inch)

Safety

Meets IEC 61010-1 +A2, CSA certified to C22.2 No.1010.1, self-certified to UL 3111

* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±5 °C from annual calibration temperature.

1 Full scale is defined as 8 vertical divisions. Magnification is used below 5 mV/div. Below 5 mV/div, full-scale is defined as 40 mV. The major scale settings are 5 mV, 10 mV,

20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V.

2 Vertical resolution for 8 bits = 0.4% of full scale, for 12 bits = 0.024% of full scale.

3 13 GHz DSP enhanced bandwidth not applicable at 5 mV/div.

4 FFT amplitude readings are affected by scope and probe bandwidth limitations and input amplifiers roll-off (e.g. 3 dB roll-off at specified bandwidth of scope/probe).

5 The FFT signal to noise ratio varies with volts/division setting, memory depth and use of time or frequency averaging.

6a Noise is the displayed noise floor. SlewRate is the displayed slew rate of the signal at the threshold crossings. Sample rate = max, sin(x)/x interpolation enabled.

6b Measurement threshold = fixed voltage at 50% level.

6c Time ranges ≤ 10 µs.

7 Values represent time error between two edges on a single channel. Standard deviation value refers to the standard deviation of 256 consecutive measurements performed

using an individual instrument. Reading is the displayed DTMA measurement value. TimeScaleAccy is the oscilloscope’s specified time scale accuracy.

8 Internal edge trigger mode. Trigger threshold = fixed voltage at 50% level. The slew rate independent value in the formula represents the traditional trigger jitter.

9 Requires Option 010 user defined function.

10 11.8 GHz analog bandwidth at 5 mV/div for DSO91304A and DSO91204A models.

11 Calculated from the bandwidth.

12 Typically triggers as low as 5 mV/div sensitivity.

13 Measurment histograms require EZJIT license