Atec Agilent-DSA90404A User Manual
Page 22
![background image](/manuals/318387/22/background.png)
22
Infiniium 90000A Series Oscilloscopes
Performance characteristics
Vertical
Input channels
Four
Analog bandwidth
(–3 dB)*
, 10
90254A
90404A
90604A
90804A
91204A
91304A
2.5 GHz
4 GHz
6 GHz
8 GHz
12 GHz
12 GHz
DSP enhanced bandwidth
3
91304A: 13-GHz real-time, user-selectable DSP enhanced bandwidth
Rise time/fall time
11
10 - 90%
20 - 80%
90254A
90404A
90604A
90804A
91204A
91304A
140 ps
105 ps
70 ps
54 ps
35 ps
32 ps
105 ps
79 ps
53 ps
38 ps
25 ps
23 ps
Input impedance
12
50 Ω, ± 3%
Sensitivity
1
1 mV/div to 1 V/div
Input coupling
DC
Vertical resolution
2
8 bits, ≥ 12 bits with averaging
Channel to channel isolation
(any two channels with
equal V/div settings)
DC to 3 GHz: 90804A/91204A/91304A: 60 dB (≥ 1000:1)
90254A/90404A/90604A: 50 dB (≥ 316:1)
3 GHz to 8 GHz: 40 dB (≥ 100:1)
8 GHz to BW: 35 dB (≥ 56:1)
DC gain accuracy*
, 1
± 2% of full scale at full resolution channel scale (± 2.5% for 5mV/div)
Maximum input voltage*
± 5 V
Offset range
Vertical sensitivity
Available offset
0 mV/div to ≥ 40 mV/div
± 0.4 V
> 40 mV/div to ≥ 75 mV/div
± 0.9 V
> 75 mV/div to ≥ 130 mV/div
± 1.6 V
> 130 mV/div to ≥ 240 mV/div
± 3.0 V
> 240 mV/div
± 4.0 V
Offset accuracy*
, 1
≤ 3.5 V: ± (2% of channel offset + 1% of full scale) + 1 mV
> 3.5 V: ± (2% of channel offset + 1% of full scale)
Dynamic range
± 4 div from center screen
DC voltage measurement
accuracy*
, 1
Dual cursor: ± [(DC gain accuracy) + (resolution)]
Single cursor: ± [(DC gain accuracy) + (offset accuracy) + (resolution/2)]
RMS noise floor (scope only)
Volts/div
5 mV
10 mV
20 mV
50 mV
100 mV
200 mV
500 mV
1 V
90254A
90404A
90604A
90804A
91204A
91304A
153 μV
199 μV
259 μV
322 μV
435 μV
467 μV
183 μV
232 μV
295 μV
358 μV
483 μV
536 μV
275 μV
342 μV
424 μV
498 μV
650 μV
758 μV
645 μV
799 μV
985 μV
1.15 mV
1.45 mV
1.73 mV
1.27 mV
1.56 mV
1.92 mV
2.22 mV
2.80 mV
3.37 mV
2.47 mV
3.03 mV
3.71 mV
4.28 mV
5.41 mV
6.58 mV
6.48 mV
8.00 mV
9.91 mV
11.5 mV
14.7 mV
17.4 mV
12.5 mV
15.6 mV
19.2 mV
22.3 mV
28.5 mV
34.1 mV
* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±5 °C from annual calibration temperature.
1 Full scale is defined as 8 vertical divisions. Magnification is used below 5 mV/div. Below 5 mV/div, full-scale is defined as 40 mV. The major scale settings are 5 mV, 10 mV, 20
mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V.
2 Vertical resolution for 8 bits = 0.4% of full scale, for 12 bits = 0.024% of full scale.
3 13 GHz DSP enhanced bandwidth not applicable at 5 mV/div.
10 11.8 GHz analog bandwidth at 5 mV/div for DSO91304A and DSO91204A models.
11 Calculated from the bandwidth.
12 Input impedance is valid when V/div scaling is adjusted to show all waveform vertical values within the scope display.