beautypg.com

Atec Agilent-DSA90404A User Manual

Page 22

background image

22

Infiniium 90000A Series Oscilloscopes

Performance characteristics

Vertical

Input channels

Four

Analog bandwidth

(–3 dB)*

, 10

90254A

90404A

90604A

90804A

91204A

91304A

2.5 GHz

4 GHz

6 GHz

8 GHz

12 GHz

12 GHz

DSP enhanced bandwidth

3

91304A: 13-GHz real-time, user-selectable DSP enhanced bandwidth

Rise time/fall time

11

10 - 90%

20 - 80%

90254A

90404A

90604A

90804A

91204A

91304A

140 ps

105 ps

70 ps

54 ps

35 ps

32 ps

105 ps

79 ps

53 ps

38 ps

25 ps

23 ps

Input impedance

12

50 Ω, ± 3%

Sensitivity

1

1 mV/div to 1 V/div

Input coupling

DC

Vertical resolution

2

8 bits, ≥ 12 bits with averaging

Channel to channel isolation

(any two channels with

equal V/div settings)

DC to 3 GHz: 90804A/91204A/91304A: 60 dB (≥ 1000:1)

90254A/90404A/90604A: 50 dB (≥ 316:1)

3 GHz to 8 GHz: 40 dB (≥ 100:1)

8 GHz to BW: 35 dB (≥ 56:1)

DC gain accuracy*

, 1

± 2% of full scale at full resolution channel scale (± 2.5% for 5mV/div)

Maximum input voltage*

± 5 V

Offset range

Vertical sensitivity

Available offset

0 mV/div to ≥ 40 mV/div

± 0.4 V

> 40 mV/div to ≥ 75 mV/div

± 0.9 V

> 75 mV/div to ≥ 130 mV/div

± 1.6 V

> 130 mV/div to ≥ 240 mV/div

± 3.0 V

> 240 mV/div

± 4.0 V

Offset accuracy*

, 1

≤ 3.5 V: ± (2% of channel offset + 1% of full scale) + 1 mV

> 3.5 V: ± (2% of channel offset + 1% of full scale)

Dynamic range

± 4 div from center screen

DC voltage measurement

accuracy*

, 1

Dual cursor: ± [(DC gain accuracy) + (resolution)]

Single cursor: ± [(DC gain accuracy) + (offset accuracy) + (resolution/2)]

RMS noise floor (scope only)

Volts/div

5 mV

10 mV

20 mV

50 mV

100 mV

200 mV

500 mV

1 V

90254A

90404A

90604A

90804A

91204A

91304A

153 μV

199 μV

259 μV

322 μV

435 μV

467 μV

183 μV

232 μV

295 μV

358 μV

483 μV

536 μV

275 μV

342 μV

424 μV

498 μV

650 μV

758 μV

645 μV

799 μV

985 μV

1.15 mV

1.45 mV

1.73 mV

1.27 mV

1.56 mV

1.92 mV

2.22 mV

2.80 mV

3.37 mV

2.47 mV

3.03 mV

3.71 mV

4.28 mV

5.41 mV

6.58 mV

6.48 mV

8.00 mV

9.91 mV

11.5 mV

14.7 mV

17.4 mV

12.5 mV

15.6 mV

19.2 mV

22.3 mV

28.5 mV

34.1 mV

* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±5 °C from annual calibration temperature.

1 Full scale is defined as 8 vertical divisions. Magnification is used below 5 mV/div. Below 5 mV/div, full-scale is defined as 40 mV. The major scale settings are 5 mV, 10 mV, 20

mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V.

2 Vertical resolution for 8 bits = 0.4% of full scale, for 12 bits = 0.024% of full scale.

3 13 GHz DSP enhanced bandwidth not applicable at 5 mV/div.

10 11.8 GHz analog bandwidth at 5 mV/div for DSO91304A and DSO91204A models.

11 Calculated from the bandwidth.

12 Input impedance is valid when V/div scaling is adjusted to show all waveform vertical values within the scope display.