beautypg.com

Atec Agilent-E5062A User Manual

Page 7

background image

7

Table 1-5

Corrected system performance with Type-N 75

device

connectors 85036E calibration kit, enhanced response calibration

Network analyzer: E5061A/E5062A, calibration kit 85036E (Type-N, 75

), calibration: enhanced response

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23°C ± 5°C with
< 1°C deviation from calibration temperature, isolation calibration not omitted

Description

Specification (dB)

300 kHz to 1.5 GHz

1.5 GHz to 3 GHz

Directivity

48

44

Source match

41

35

Load match

15

15

Reflection tracking

±0.010

±0.019

Transmission tracking

±0.015

±0.029

Transmission uncertainty (specification)

Magnitude

Phase

Reflection uncertainty (specification)

Magnitude

Phase