Table 1-20

Typical cycle time for measurement completion

1

(ms) (Display update: off)

Number of points

51

201

401

1601

Start 1 GHz, stop 1.2 GHz, 30 kHz IF bandwidth

Uncorrected

8

19

33

117

2-port cal

14

35

63

230

Start 300 kHz, stop 1.5 GHz, 30 kHz IF bandwidth

Uncorrected

15

25

39

123

2-port cal

27

48

75

243

Start 300 kHz, stop 3 GHz, 30 kHz IF bandwidth

Uncorrected

17

28

41

125

2-port cal

31

53

80

247

Measurement throughput summary

1. Typical performance.

Table 1-21

Typical cycle time for measurement completion

1

(ms) (Display update: on)

Number of points

51

201

401

1601

Start 1 GHz, stop 1.2 GHz, 30 kHz IF bandwidth

Uncorrected

59

68

83

172

2-port cal

85

103

131

304

Start 300 kHz, stop 1.5 GHz, 30 kHz IF bandwidth

Uncorrected

64

74

89

178

2-port cal

95

116

144

317

Start 300 kHz, stop 3 GHz, 30 kHz IF bandwidth

Uncorrected

66

78

91

180

2-port cal

98

121

148

322