beautypg.com

Atec Agilent-E5062A User Manual

Page 5

background image

5

Table 1-3

Corrected system performance with Type-N device connectors,
85032F calibration kit, enhanced response calibration

Network analyzer: E5061A/E5062A, calibration kit: 85032F (Type-N, 50

) calibration: enhanced

response

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23°C ± 5°C with
< 1°C deviation from calibration temperature, isolation calibration not omitted

Description

Specification (dB)

300 kHz to 1.5 GHz

1.5 GHz to 3 GHz

Directivity

49

46

Source match

41

40

Load match

15

15

Reflection tracking

±0.011

±0.021

Transmission tracking

±0.015

±0.018

Transmission uncertainty (specification)

Magnitude

Phase

Reflection uncertainty (specification)

Magnitude

Phase