beautypg.com

Atec Agilent-E5062A User Manual

Page 19

background image

19

Table 1-22

Data transfer time

1

(ms)

Number of points

51

201

401

1601

SCPI over GPIB

2

REAL 64

6

14

25

89

ASCII

51

193

383

1522

SCPI over 100 Mbps LAN (telnet)

2

REAL 64

3

3

4

6

ASCII

92

354

510

2040

SCPI over 100 Mbps LAN (SICL-LAN)

2

REAL 64

7

7

8

12

ASCII

9

21

34

127

COM (program executed in the analyzer)

2

Variant type

2

2

2

2

1. Typical performance.
2. Measured using a VEE 6.01 program running on a 500 MHz Pentium

®

III Dell Optiplex, Transferred

complex S

11

data, using :CALC:DATA?SDATA.