Measurement capabilities

Source control

Number of measurement channels

Up to 4 independent measurement channels. A measurement channel is coupled
to stimulus response settings including frequency, IF bandwidth, power level,
and number of points.

Number of display windows

Each measurement channel has a display window. Up to 4 display windows (channels)
can be displayed.

Number of traces

4 data traces and 4 memory traces per channel

Measurement choices

Option E5061A/E5062A-150/175: S

11

, S

21

Option E5061A/E5062A-250/275: S

11

, S

21

, S

12

, S

22

Measurement parameter conversion

Available to convert S-parameters into reflection impedance, transmission impedance,
reflection admittance, transmission admittance, and 1/S.

Data formats

Log magnitude, linear magnitude, phase, expanded phase, positive phase,
group delay, SWR, real, imaginary, Smith chart, polar.

Data markers

10 independent markers per trace. Reference marker available
for delta marker operation. Smith chart format includes 5 marker formats:
linear magnitude/phase, log magnitude/phase, real/imaginary,
R + jX, and G + jB. Polar chart format includes 3 marker formats:
linear magnitude/phase, log magnitude/phase, and real/imaginary.

Marker functions

Marker search

Max value, min value, multi-peak, multi-target, peak, peak left, peak right, target,
target left, target right, and width parameters with user-defined bandwidth values.

Marker-to functions

Set start, stop, center to active marker stimulus value; set reference to active
marker response value; set electrical delay to group delay at active marker.

Search range

User definable.

Tracking

Performs marker search continuously or on demand.

Fault location functions (Option E5061A/E5062A-100)

Transformation to distance and time domain

Selectable transformation type from bandpass, lowpass impulse, lowpass step.
Selectable window from maximum, normal and minimum.

Measured number of points per sweep

User definable from 2 to 1601.

Sweep type

Linear sweep, segment sweep, log sweep and power sweep.

Segment sweep

Define independent sweep segments. Set number of points, test port power levels,
IF bandwidth, delay time, sweep time and sweep mode independently for each segment.

Sweep trigger

Set to continuous, hold, or single, sweep with internal, external, manual,
or bus trigger.

Power

Set source power from -5 dBm (-45 dBm for option E5061A/E5062A-1E1/250/275)
to 10 dBm. The power slope function compensates source power level error.