beautypg.com

Corrected system performance – Atec Agilent-E5062A User Manual

Page 3

background image

3

Corrected system performance

The specifications in this section apply for measurements made with the
Agilent E5061A/E5062A network analyzer with the following conditions:

• No averaging applied to data

• Environmental temperature of 23°C ± 5°C, with less than 1°C deviation

from the calibration temperature

• Response and isolation calibration not omitted

Table 1-1

System dynamic range

1

Description

Specification

Supplemental information

System dynamic range

300 kHz to 3 GHz, IF bandwidth = 3 kHz

90 dB

95 dB

300 kHz to 3 GHz, IF bandwidth = 10 Hz

115 dB

120 dB

1. The test port dynamic range is calculated as the difference between the test port rms noise floor

and the source maximum output power. The effective dynamic range must take measurement
uncertainty and interfering signals into account.