beautypg.com

Atec Agilent-E5062A User Manual

Page 6

background image

6

Table 1-4

Corrected system performance with Type-N 75

device

connectors 85036E calibration kit, full 2-port calibration

Network analyzer: E5061A/E5062A, calibration kit: 85036E (Type-N, 75

), calibration: full 2-port

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23°C ± 5°C with
< 1°C deviation from calibration temperature, isolation calibration not omitted

Description

Specification (dB)

300 kHz to 1.5 GHz

1.5 GHz to 3 GHz

Directivity

48

44

Source match

41

35

Load match

48

44

Reflection tracking

±0.010

±0.019

Transmission tracking

±0.015

±0.029

Transmission uncertainty (specification)

Magnitude

Phase

Reflection uncertainty (specification)

Magnitude

Phase