14
Table 1-17
Rear panel information
Description
Supplemental information
External trigger connector
Type
BNC, female
Input level
LOW threshold voltage: 0.5 V
HIGH threshold voltage: 2.1 V
Input level range: 0 to +5 V
Pulse width
≥ 20 µsec, typical
Polarity
Negative (downward) only
External reference signal input connector
Type
BNC, female
Input frequency
10 MHz ±10 ppm, typical
Input level
0 dBm ±3 dB, typical
Internal reference signal output connector
Type
BNC, female
Output frequency
10 MHz ±10 ppm, typical
Signal type
Sine wave, typical
Output level
0 dBm ±3 dB into 50
Ω, typical
Output impedance
50
Ω, nominal
VGA video output
15-pin mini D-Sub; female; drives VGA compatible monitors
GPIB
24-pin D-Sub (type D-24), female; compatible with IEEE-488
Parallel port
36-pin D-Sub (type 1284-C), female; provides connection to printers, or multiport
test set
USB port
Universal serial bus jack, type A configuration (4 contacts inline, contact 1 on left);
female; provides connection to printer, ECal module, USB/GPIB interface
Contact 1
Vcc: 4.75 to 5.25 VDC, 500 mA, maximum
Contact 2
-Data
Contact 3
+Data
Contact 4
Ground
LAN
10/100 BaseT Ethernet, 8-pin configuration; auto selects between the two data rates
Handler I/O port
36-pin D-sub, female; provides connection to handler system
Line power
1
Frequency
47 Hz to 63 Hz
Voltage
90 to 132 VAC, or 198 to 264 VAC (automatically switched)
VA max
350 VA max.
1. A third-wire ground is required.