beautypg.com

Atec Agilent-E5062A User Manual

Page 14

background image

14

Table 1-17

Rear panel information

Description

Supplemental information

External trigger connector

Type

BNC, female

Input level

LOW threshold voltage: 0.5 V
HIGH threshold voltage: 2.1 V
Input level range: 0 to +5 V

Pulse width

≥ 20 µsec, typical

Polarity

Negative (downward) only

External reference signal input connector

Type

BNC, female

Input frequency

10 MHz ±10 ppm, typical

Input level

0 dBm ±3 dB, typical

Internal reference signal output connector

Type

BNC, female

Output frequency

10 MHz ±10 ppm, typical

Signal type

Sine wave, typical

Output level

0 dBm ±3 dB into 50

Ω, typical

Output impedance

50

Ω, nominal

VGA video output

15-pin mini D-Sub; female; drives VGA compatible monitors

GPIB

24-pin D-Sub (type D-24), female; compatible with IEEE-488

Parallel port

36-pin D-Sub (type 1284-C), female; provides connection to printers, or multiport
test set

USB port

Universal serial bus jack, type A configuration (4 contacts inline, contact 1 on left);
female; provides connection to printer, ECal module, USB/GPIB interface

Contact 1

Vcc: 4.75 to 5.25 VDC, 500 mA, maximum

Contact 2

-Data

Contact 3

+Data

Contact 4

Ground

LAN

10/100 BaseT Ethernet, 8-pin configuration; auto selects between the two data rates

Handler I/O port

36-pin D-sub, female; provides connection to handler system

Line power

1

Frequency

47 Hz to 63 Hz

Voltage

90 to 132 VAC, or 198 to 264 VAC (automatically switched)

VA max

350 VA max.

1. A third-wire ground is required.