Power meter (option 0029) – Atec Anritsu-MT8221B User Manual
Page 6
Page 6 of 24
BTS Master™ MT8221B Base Station Analyzer Specifications
Power Meter (Option 0029)
Frequency
Center/Start/Stop, Span, Frequency Step, Signal Standard, Channel #, Full Band
Amplitude
Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale
Average
Acquisition Fast/Med/Slow, # of Running Averages
Limits
Limit On/Off, Limit Upper/Lower
Frequency Range
10 MHz to 7.1 GHz
Span
1 kHz to 100 MHz
Display Range
-140 dBm to +30 dBm, ≤40 dB span
Measurement Range
-120 dBm to +30 dBm
Offset Range
0 to +100 dB
VSWR
1.5:1 typical
Maximum Power
+30 dBm without attenuator
Accuracy
Same as Spectrum Analyzer
Application Options
Impedance (50 Ω, 75 Ω, Other)
High Accuracy Power Meter (Option 0019) (Requires external USB Power Sensor(s))
Amplitude
Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale
Average
# of Running Averages, Max Hold
Zero/Cal
Zero On/Off, Cal Factor (Center Frequency, Signal Standard)
Limits
Limit On/Off, Limit Upper/Lower
Power Sensor Model
PSN50
MA24104A
MA24106A
MA24108A
MA24118A
Description
High Accuracy
RF Power Sensor
Inline High
Power Sensor
High Accuracy
RF Power Sensor
Microwave USB
Power Sensor
Microwave USB
Power Sensor
Frequency Range
50 MHz to 6 GHz
600 MHz to 4 GHz
50 MHz to 6 GHz
10 MHz to 8 GHz
10 MHz to 18 GHz
Connector
Type N(m), 50 Ω
Type N(m), 50 Ω
Type N(m), 50 Ω
Type N(m), 50 Ω
Type N(m), 50 Ω
Dynamic Range
-30 to +20 dBm
(.001 to 100 mW)
+3 to +51.76 dBm
(2 mW to 150 W)
-40 to +23 dBm
(0.1 µW to 200 mW)
-40 to +20 dBm
(0.1 µW to 100 mW)
-40 to +20 dBm
(0.1 µW to 100 mW)
VBW
100 Hz
100 Hz
100 Hz
50 kHz
50 kHz
Measurand
True-RMS
True-RMS
True-RMS
True-RMS,
Slot Power,
Burst Average
Power
True-RMS,
Slot Power,
Burst Average
Power
Measurement Uncertainty
±0.16 dB
1
±0.17 dB
2
±0.16 dB
1
±0.18 dB
3
±0.18 dB
3
Datasheet
(for complete specifications)
11410-00414
11410-00483
11410-00424
11410-00504
11410-00504
Notes:
1) Total RSS measurement uncertainty (0 ºC to 50 ºC) for power measurements of a CW signal greater than -20 dBm with
zero mismatch errors.
2) Expanded uncertainty with K=2 for power measurements of a CW signal greater than +20 dBm with a matched load.
Measurement results referenced to the input side of the sensor.
3) Expanded uncertainty with K=2 for power measurements of a CW signal greater than-20 dBm with zero mismatch errors.