Operation

Theoretical Sample of Charging Field

N

UMBER

D

IGITS

Number Digits allows the selection of the number of digits, either three or four,

to be displayed.

AVG

M

ODE

AVG Mode allows the selection between MEAN and RMS (root mean square).

Mean is the average of the observed values from the beginning of the

averaging interval.

RMS refers to the square root of the mean of the squared value of a

function. The RMS average field intensity value is proportional to the

power density of the field. When logging in power density units, the

RMS and mean values are equal.

For more information on averaging see Averaging on page 38.

Data

Options under the data menu include: Log View, Log Upload, and Log Delete.

L

OG

V

IEW

Log View displays the logged data files stored in the unit.