beautypg.com

ETS-Lindgren HI-2200 RF Survey Meter User Manual

Page 31

background image

Operation

| 31

In spatial averaging, a series of measurements are taken during a

relatively short period of time (usually not more than two to three

minutes). At the end of the period, the average value is the answer.

The measurement is used to monitor compliance with standards,

based on the whole-body average maximum permissible exposures

(MPE). The average field is measured over the space occupied by the

subject’s body.

For more information about averaging see Averaging on page 38.

A

LARM

P

OINT

Alarm Point allows the user to set a field level above which an alarm will

indicate/sound. The range is 0.00 to 1000 units. Audio must be set to Alarm for

an audible tone to sound when the alarm level is reached. If the alarm level is set

higher than the range of the probe, the alarm will not indicate/sound. Setting the

alarm level to 0.00 will turn the alarm off.

A

UDIO

Audio is an audio annunciator that is provided to signal various functions on the

HI-2200. Options include None, Alarm, Cadence, and Both.

None turns off all sounds excluding the startup tone.

Alarm causes the unit to emit a tone when the alarm level set under

the Alarm Point option is exceeded.

Cadence enables a pulsing tone to assist in measuring consistently

over a distance. This tone sounds every second during logging in

continuous or spatial log mode.

Both enables the alarm point tone and the cadence tone at the same

time. In addition to the pulse tone, a continuous tone will sound when

the alarm level is reached