beautypg.com

0 specifications – ETS-Lindgren HI-2200 RF Survey Meter User Manual

Page 15

background image

Specifications

| 15

3.0 Specifications

Meter Specifications

Internal Memory:

>65,000 data points

User Selectable Units

(Probe Dependent):

V/m,

μW/cm

2

, mW/cm

2

, W/m

2

A/m,

mA/m

Battery:

Four AA

(1.2–1.5 V Nominal Alkaline or

Nickel-Metal Hydride or

Nickel-Cadmium)

Operating Life:

Approximately 20 hours

(with new alkaline batteries)

Operating Temperature:

-10

o

C to 50

o

C

Connector:

RS-232 I/O serial interface six-pin

mini DIN

Probe Specifications

Probes must be calibrated or matched prior to using them with the

HI-2200. For more information, see Matched Probes on page 28.

The following tables represent probes that are designed to be used with the

HI-2200 RF Survey Meter.

M

ODEL

E100

E-F

IELD

P

ROBE

Model E100 E-Field Probe

Sensor Type:

Electric Field

Detection:

Isotropic (Total Field Indication)

Frequency Range:

100 kHz – 5 GHz

Frequency Response:

± 1 dB (1 MHz to 4 GHz)

Dynamic Range:

0.3 V/m to 800 V/m (Single Range)