Aed procedure, Aed procedure -6 – Physio-Control LIFEPAK 20e User Manual
Page 76
Therapy
4-6
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
When the CPR metronome is set to
ON
in the setup options (refer to
), the CPR
metronome icon
is displayed and the metronome sounds automatically during CPR times.
You cannot enable or disable the CPR metronome while in AED mode. Refer to
, for more information.
AED Procedure
The following descriptions of voice prompts and messages are based on the factory default
settings for AED mode. The default settings are consistent with 2010 American Heart
Association (AHA) and European Resuscitation Council (ERC) guidelines. Changing the setup
options may result in different AED behavior. Refer to
1 Verify that the patient is in cardiopulmonary arrest (unresponsive, not breathing normally, and
showing no signs of circulation.)
2 Press
ON
.
3 Prepare the patient for electrode placement (refer to
"Paddles Monitoring Procedure"
on
4 Connect the therapy electrodes to the therapy cable, and confirm cable connection to the
defibrillator.
5 Apply the electrodes to the patient's chest in the anterior-lateral position (refer to
).
When the ECG waveform is set to
OFF
in
the setup options, the messages and
prompts fill the screen as shown in the
screen to the left.
The
CONNECT ELECTRODES
message
and voice prompt occur until the patient
is connected to the AED.