Table 2-2 – Physio-Control LIFEPAK 20e User Manual
Page 44
Basic Orientation
2-22
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
Table 2-2
Battery Status Indicator
Battery
Indicator
Battery Status
Messages and Tones
Battery
Capacity
(percentage)
Four green bars
None
76–100
Three green bars
None
51–75
Two green bars
None
26–50
One green bar
None
16–25
One yellow bar
LOW BATTERY
Three beeps (one time). Message
remains in message area, cycling with
others if necessary.
11–15
One flashing red bar
LOW BATTERY: CONNECT TO AC POWER
Three beeps (every 20 seconds).
Message remains in message area,
cycling with others if necessary.
0–10
It is important to note that both charge (voltage) and capacity are factors contributing to the
battery status indicator’s display. For example, the battery status indicator displays two green
bars when a new battery has 100% capacity and has been charged to 50%. However, an
older battery that is 100% charged and only has 50% capacity also displays two green bars.
The following demonstrates this example.
This battery may either be low in capacity or low in charge as follows:
• For a new battery with 100% capacity, this symbol indicates 50% charged (low in
charge).
• For an older battery with 50% capacity, this symbol indicates 100% charged (low
in capacity).