Physio-Control LIFEPAK 20e User Manual
Page 156
Specifications and Performance Characteristics
A-4
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
CO
2
CO
2
Range
0 to 99 mmHg (0 to 13.2 kPa)
Units: mmHg, %, or kPa
CO
2
Accuracy
CO
2
partial pressure at sea
level:
Accuracy:
(0–80 bpm)*
0 to 38 mmHg
(0 to 5.1 kPa)
±2 mmHg
(0.27 kPa)
39 to 99 mmHg
(5.2 to 13.2 kPa)
±5% of reading + 0.8% for
every 1 mmHg (0.13 kPa)
above 38 mmHg (5.1 kPa)
(>80 bpm)*
0 to 18 mmHg
(0 to 2.4 kPa)
±2 mmHg
(0.27 kPa)
19 to 99 mmHg
(2.53 to 13.2 kPa)
±4 mmHg (0.54 kPa) or
±12% of reading, whichever
is higher
*For RR > 60 bpm, to achieve specified CO
2
accuracy, the
Microstream
®
Filterline
®
H Set for infants must be used.
Drift of Measurement
Accuracy
No drift in accuracy for at least 6 hours
Respiration Rate
Accuracy
0 to 70 bpm: ±1 bpm
71 to 99 bpm: ±2 bpm
Respiration Rate
Range
0 to 99 breaths/minute
Flow Rate
42.5 to 65 ml/min (measured by volume)
Rise Time
190 msec
Response Time
4.5 seconds, maximum (includes delay time and rise time)
Initialization Time
30 seconds (typical), 10–180 seconds
Ambient Pressure
Automatically compensated internally
Waveform Scale
Factors
Autoscale, 0–20 mmHg (0–4 Vol%), 0–50 mmHg (0–7 Vol%),
0–100 mmHg (0–14 Vol%)
PRINTER
Prints continuous strips of the displayed patient information
Paper Size
50 mm (2.0 in.)
Print Speed
Continuous ECG 25 mm/sec
±
5% (measured in accordance
with AAMI EC-11 (1991), 4.2.5.2)
25 mm/sec print speed for CODE SUMMARY Reports
Delay
8 seconds
Autoprint
Waveform events print automatically (user configurable)