Physio-Control LIFEPAK 20e User Manual
Page 153
Appendix A
Specifications and Performance Characteristics
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
A-1
©2006-2013 Physio-Control, Inc.
All specifications are at 20°C (68°F) unless otherwise stated.
GENERAL
The LIFEPAK 20e defibrillator/monitor has seven main operating modes:
Manual Mode
Provides normal operating capability for ALS users
AED Mode
Provides normal operating capability for BLS users
Archive Mode
Allows operator to print, edit or delete previous patient
records
Setup Mode
Allows operator to configure the instrument
Service Mode
Allows operator to execute device diagnostic tests and
calibrations
Inservice Mode
Provides simulated waveforms for demonstration
purposes
Auto Test
Mode
Provides daily automatic tests of critical circuits
POWER
AC Powered
100-120 VAC 50/60 HZ, 220-240 VAC 50/60 HZ. Total
power draw less than 120 Volt-Amperes (VA), or less
than 150 VA with the CodeManagement Module
installed.
Operating Time
A new, fully charged, internal backup battery provides a minimum of the following prior to
shutdown:
---total--
--after low battery--
Monitoring plus SpO
2
(minutes)
210
5
Defibrillation (360 J
discharges):
140
3
Monitoring Plus Pacing
(minutes at 100 mA, 60 ppm)
Plus SpO
2
:
110
2
Battery Charging Time
Battery charges while defibrillator operates from AC
power. Typical battery charge time of less than four
hours when defibrillator is powered off and AC power is
applied.
Low Battery Indication
Low battery message in status area and warning tone
sounds.
Battery Status Indicator
Indicates available battery capacity for defibrillator.
A Service Indicator
When Error Detected.