Data transfer from truecpr device, Data transfer from tru, Figure 6-3 – Physio-Control LIFEPAK 20e User Manual
Page 121: Represents
![background image](/manuals/313536/121/background.png)
6
D
ata Mana
g
e
ment
Data Management
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
6-13
©2006-2013 Physio-Control, Inc.
Figure 6-3
IrDA Connections
DATA TRANSFER FROM TrueCPR DEVICE
Note: The TrueCPR device may not be available in all countries. Contact your local
Physio-Control representative for more information.
You can transfer data from the Physio-Control TrueCPR device to Physio-Control post-event
review tools using the TrueCPR device port on the CodeManagement Module. To transfer data:
1 Confirm the LIFEPAK 20e defibrillator/monitor is either turned off or in archives mode.
2 Connect an approved USB cable between the TrueCPR device and the CodeManagement
Module.
Note: Use only a USB cable that meets the following specifications: USB 2.0 A-Male to
Mini-B 5-pin Male, 28/24 AWB cable with ferrite core (gold plated), 1 m (3 ft).
3 Turn on the TrueCPR device and follow the instructions in the TrueCPR device Instructions
for Use.
Notes:
• It may take up to one minute for the TrueCPR device to display the Data Transfer screen.
• Data transfer takes approximately 3–5 minutes.
• If you attempt to transmit data from the LIFEPAK 20e defibrillator/monitor and the TrueCPR
device simultaneously, the data from the defibrillator will be transmitted first.
IrDA
Adapter/
Computer
Maximum Distance: 1.0 m (3.28 feet)
Defibrillator