Physio-Control LIFEPAK 20e User Manual
Page 155
Appendix A
Specifications and Performance Characteristics
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
A-3
©2006-2013 Physio-Control, Inc.
Lead Selection
Leads I, II, III (3-wire ECG cable)
Leads I, II, III, AVR, AVL, AVF, and C acquired
simultaneously (5-wire ECG cable)
ECG Size
4, 3, 2.5, 2, 1.5, 1, 0.5, 0.25 cm/mV
Heart Rate Display
20–300 bpm digital display
Out of range indication: Display symbol "---"
Heart symbol flashes for each QRS detection
Continuous Patient
Surveillance System
(CPSS)
In AED mode, while Shock Advisory System is not active,
CPSS monitors the patient, via QUIK-COMBO paddles or
Lead II ECG, for potentially shockable rhythms.
Voice Prompts
Used for selected warnings and alarms (configurable on/off).
Analog ECG Output
1 V/mV x 1.0 gain <35 msec delay
Common Mode
Rejection
90 dB at 50/60 Hz
SpO
2
Masimo
®
sensors
Saturation range
1 to 100%
Saturation Accuracy
70–100% (0–69% unspecified)
Adults/Pediatrics
±2 digits (during no motion conditions)
±3 digits (during motion conditions)
Neonates
±3 digits (during no motion conditions)
±3 digits (during motion conditions)
Dynamic signal
strength bar graph
Pulse tone at the onset
of the pleth waveform
SpO
2
Update
Averaging Rate
User selectable 4, 8, 12 or 16 seconds
SpO
2
measurement
Functional SpO
2
values are displayed and stored
Pulse rate range
25 to 240 pulses per minute
Pulse rate accuracy
Adults/Pediatrics/
Neonates
±3 digits (during no motion conditions)
±5 digits (during motion conditions)
SpO
2
waveform with
autogain control
Alarms
Quick Set
Activates alarms for all parameters
VF/VT Alarm
Activates continuous CPSS monitoring in Manual Mode