Physio-Control LIFEPAK 20e User Manual
Page 164
Specifications and Performance Characteristics
A-12
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
CHARGE TIME
AC Operation Only:
Maximum Time from Charge to Shock Ready
Voltage
Manual Mode
AED Mode
100–120V
360 J
7 seconds
200 J
5 seconds
220–240V
360 J
7 seconds
200 J
5 seconds
90V (90% of Nominal -100)
360 J
7 seconds
200 J
5 seconds
198V (90% of Nominal -220) 360 J
7 seconds
200 J
5 seconds
Maximum Time from Initiation of Analysis to Shock Ready (AED only)
Voltage
AED Mode
100–120V
200 J
12 seconds
220–240V
200 J
12 seconds
90V (90% of Nominal -100)
200 J
12 seconds
198V (90% of Nominal -220)
200 J
12 seconds
Maximum Time from Power-on to Shock Ready
Voltage
Manual Mode
AED Mode
90V (90% of Nominal -100)
360 J
11 seconds
200 J
16 seconds
198V (90% of Nominal -220) 360 J
11 seconds
200 J
16 seconds
Battery-only Operation:
Maximum Time from Charge to Shock Ready
Battery
Manual Mode
AED Mode
Fully charged
360 J
7 seconds
200 J
5 seconds
Fully charged battery,
followed with 15 full energy
discharges
360 J
7 seconds
200 J
5 seconds
Maximum Time from Initiation of Analysis to Shock Ready (AED only)
Battery
AED Mode
Fully charged
200 J
12 seconds
Fully charged battery, followed with
15 full energy discharges
200 J
12 seconds
Maximum Time from Power-on to Shock Ready
Battery
Manual Mode
AED Mode
Fully charged battery,
followed with 15 full energy
discharges
360 J
11 seconds
200 J
16 seconds