background image

Polychromix DTS TM Near-Infrared Spectrometer

A-1

Operator’s Manual

Document Number 04-91-0201-00 Rev A

Appendix A Specifications

Specifications

DTS

1700

DTS

2500

DTS

NB

Spectral Range

0.9-1.7 microns

1.7-2.5 microns

1.1-1.35 microns

Absolute Wavelength
Accuracy

±0.6nm fixed temperature (with parabolic fit)

Resolution

Pixel Spacing

8 nm

8 nm

2 nm

Optical Resolution

12 nm

22 nm

4 nm

Stray Light

<0.01% transmission of 10mm water at 1405 nm using Halogen lamp &
850 nm filter

Dynamic Range

10,000 single acquisition with optimum light levels

Photometric Stability

<0.05% over 6 hrs.

Measurement time

<10 msec per spectral mask: ,<1 sec. per spectrum

Input fiber

for optimum performance internal slit needs

500 um diameter low OH

fiber with SMA-905 connector.

Module Size
(w/electronics)

105 x 85 x 145mm (4.2x3.3x5.7”)

Weight

1.8 Kg (4 lbs.)

Hardware Interface

mini USB for both power and communications.

A/D Converter

24 bit

Power consumption

0.75W

1.25W

0.75W

Supply Requirement

5 Volts DC, 400mA
400mA

Detector

Single InGaAs

Single InGaAs with 2
stage cooler

Single InGaAs

Software included:

SpectralCode

Contact Polychromix for: “my instrument” or Labview

driver

Operating Systems

Windows Me, 2000, & XP

Operating Environment
Range

5 to 40°C and 0 to 80% RH

Storage Temp Range

-40 to +60°C