beautypg.com

Ocean Optics DTS User Manual

Page 34

background image

Chapter 4

4-10

Polychromix DTS

TM

Near-Infrared Spectrometer

Operator’s Manual

Document Number 04-91-0201-00 Rev A

If the user now performs a transmittance scan, that scan will be placed in

register 0 and can be moved with the

key to register 2 (Figure 4-13).

Figure 4-13: The Plot Legend Dialog Box after Collecting a Transmittance Scan
and moving it.

The color of the traffic light in the PlotLegend dialog box indicates the display
status of the register.

Red - The contents of the register are not displayed.

Yellow – The contents of the register can be displayed if the appropriate
spectral type is active (e.g. a reflectance scan cannot be displayed if the
absorbance mode is active, but could be displayed if the active mode was
changed to reflectance).

Green - The contents of the register are displayed.

To change the display state:

To display a register that is not currently displayed, click on the traffic light

until it turns green.

To turn off a register that is currently displayed, click on the traffic light

until it turns red.

The highlighted scan in the Plot Legend dialog box (which is indicated with a
gray background) can be renamed as desired by typing in the desired name.