Maintenance and Troubleshooting

Polychromix DTS

TM

Near-Infrared Spectrometer

7-3

Operator’s Manual

Document Number 04-91-0201-00 Rev A

7.3.2

If the Spectrum is Noisy

Cause

Comment

Solution

Lamp is weak

The lamp has a finite
lifetime

Replace the lamp

The absorbance is weak or
% T is very high (noisy
spectrum)

The sample is too dilute or
too thin

Prepare a more
concentrated sample

Use Spectrum Averaging
to reduce noise (Section
5.4)

The absorbance is too
strong

The sample is too
concentrated or thick

Dilute the sample or
reduce its thickness

Cable is loose

Check all cables for good
contact

Fiber optic cable is
defective

Replace fiber optic cable

Transient internal
spectrometer fault

Remove USB cable and
re-insert

Defect in Spectrometer

Contact Polychromix or its
local representative

7.3.3

There is no Signal

Cause

Comment

Solution

USB cable is loose

Check cables for good
contact

Light not on.

Check light

Sample not appropriate

Check light meter

Transient internal
spectrometer fault

Remove USB cable and
re-insert

Defect in Spectrometer

Contact Polychromix or its
local representative