beautypg.com

Ocean Optics DTS User Manual

Page 50

background image

Chapter 6

6-4

Polychromix DTS

TM

Near-Infrared Spectrometer

Operator’s Manual

Document Number 04-91-0201-00 Rev A

Table 6-1: Sample Related Issues

Problem

Cause

Solution

The absorbance is
weak or % T is very
high (noisy spectrum)

The sample is too dilute
or too thin

Prepare a more
concentrated sample

Use Spectrum
Averaging to reduce
noise (Section 6.4)

The absorbance is too
strong or % T is near
zero

The sample is too
concentrated or thick

Dilute the sample or
use a thinner cuvette

Intensity of peaks
change

Sample Decomposition

Use a thermostated
cuvette holder

Spectrum does not
compare to spectrum in
another solvent

Interaction between
sample and solvent

Replace solvent

6.2.6

Quantitative Measurements

The transmission (T) of a sample is defined by equation 6-1.

T = P/P

o

6-1

where: P is the power of the light that is transmitted by the sample

P

o

is the power of the light incident on the sample

The absorbance (A) of a sample is the logarithm (base 10) of the reciprocal of
the transmittance (equation 6-2).

A = -log T = log (1/T)

6-2

Beer’s Law (equation 6-3), can be used to determine the concentration of the
compound of interest.

A = e x c x l

6-3

where: A is the absorbance

e is the molar absorbance coefficient, which is a constant for a

compound in a given solvent at a given temperature
c is the concentration of the compound
l is the path length.