The Signal Intensity is not Constant

Cause

Comment

Solution

Fiber optic cable has
shifted in position

Tape fiber optic cable to
bench

Lamp is not stable

Defective lamp

Replace the lamp

Sample is not
homogeneous
(Cuvette)

Mix sample and filter if
necessary to remove
particulate matter

Sample is not
homogeneous
(Integrating Sphere)

Move integrating sphere to
a different region of the
sample

Sample is decomposing

Replace sample

Cuvette is scratched

Always place cuvette in
sample holder with same
plane facing the source

Replace cuvette

Defect in Spectrometer

Contact Polychromix or its
representative

7.3.5

The Wavelength for a Peak has Shifted

Cause

Comment

Solution

The background spectrum
is not valid

Run a new background
Spectrum

Sample is not
homogeneous

Find another region of the
sample

Defect in Spectrometer

Contact Polychromix or its
local representative