6 set-up, 1 adjustment structure – VEGA VEGAPULS 56V User Manual
Page 48
48
VEGAPULS 56V
Set-up
Adjustment with the adjustment module
MINICOM
With the adjustment module MINICOM you
adjust directly in the sensor or in the external
indicating instrument VEGADIS 50. The ad-
justment module MINICOM enables with the
6-key adjustment field with text display the
parameter adjustment of the sensor in the
same function volume as with the adjustment
program VVO or the signal conditioning in-
strument VEGAMET, however not the con-
figuration of the measuring system.
The adjustment of the signal conditioning
instrument is only possible with the adjust-
ment program VVO or the 6-key adjustment
field on the signal conditioning instrument.
Independent whether you set-up a measur-
ing system (unit of sensor and signal condi-
tioning instrument VEGAMET or processing
system VEGALOG) with the adjustment soft-
ware VVO or with the signal conditioning
instrument, the adjustment procedure is
always the same:
- first of all configure a measuring system in
the menu "Configuration“ and then
- carry out the parameter adjustment of the
sensor in the menu "Instrument data“.
Before starting with the set-up
:
Do not be confused by the many pictures,
adjustment steps and menus on the following
pages. Carry out the set-up with the PC
step-by-step and soon you will no more
require the following pages.
6 Set-up
6.1 Adjustment structure
VEGAPULS 56 radar sensors can be ad-
justed with
- PC (adjustment program VVO),
- detachable adjustment module MINICOM
or
- signal conditioning instrument VEGAMET.
The adjustment must only be carried out with
one adjustment medium at the same time.
Adjustment with PC
The PC with adjustment program VVO (VEGA
Visual Operating) can be connected to the:
- sensor
- signal line
- signal conditioning instrument VEGAMET
514V/515V
- processing system VEGALOG 571
With the adjustment program VVO (VEGA
Visual Operating) on the PC you can adjust
the radar sensors in a very comfortable way.
The PC communicates via the interface con-
verter VEGACONNECT␣ 2 with the sensor and
the signal conditioning instrument or with the
standard RS␣ 232-interface cable, with the
processing system VEGALOG and all con-
nected sensors. Therefore a digital adjust-
ment signal is superimposed to the signal
and supply line of the sensors. The adjust-
ment can be therefore made directly on the
sensor, on any individual position of the sig-
nal line or on the processing system VEGA-
MET or VEGALOG.
Adjustment with the signal conditioning
instrument VEGAMET
Like with the adjustment program VVO, sen-
sor and signal conditioning instrument
VEGAMET can be adjusted with the 6-key
adjustment field on the signal conditioning
instrument. The adjustment is possible in the
same function volume than with the adjust-
ment program VVO on the PC.