10 returning to a weighing mode, 11 returning to a preset apw, 12 display hold modes – Ohaus CD-11 Indicator Manual multi User Manual
Page 35: En-29 cd-11 indicator, 1 manual display hold (dhman)
EN-29
CD-11 Indicator
4.10 Returning to a Weighing Mode
Press the PRINT/
Units button until the display indicates the desired measuring
unit either kg, lb, g or oz.
4.11 Returning to a Preset APW
If the APW has been calculated previously, the Indicator stores the value in
memory. Proceed as follows to use a previously set APW:
Press and hold the PRINT/
Units button until PCS cursor is displayed.
Place samples on the platform. The display indicates the number of pieces
based on the previously entered data. Sample shown at right indicates 100
pieces.
Stable
Gross
Brutto
CAUTION
WHEN POWER IS TURNED OFF, APW
WILL ALWAYS RETURN TO THE PREVIOUSLY STORED APW.
PCS Gross
Brutto
4.12 Display Hold Modes
The Display Hold mode is enabled by entering the Alternate mode sub-menu in
the rEAd menu. The Alternate mode selections are: ALtOFF, ALt PC, ALt dH.
To enable the Display Hold mode, select ALt dH.
When ALt dH (display hold) is selected, the following choices are available:
dHMAn (manual display hold), dHSEM (semi-automatic display hold), and
dHAuto (automatic display hold). dHMan is the default setting. These choices
are displayed only if ALt dH is selected.
The Display Hold mode allows the maximum stable weight value to be held on
the display and sent to a peripheral device prior to being cleared.
The Display Hold types operate as described below:
4.12.1 Manual Display Hold (dHMAn)
To activate the Manual Display Hold mode, press and hold the PRINT/
Units
button until the LCD cursor lights above the Display Hold text on the panel.
The cursor will blink to indicate that the scale is ready. Display shown at right
indicates an empty platform.