beautypg.com

En-12 cd-11 indicator, 4 setup menu (cont.) – Ohaus CD-11 Indicator Manual multi User Manual

Page 18

background image

EN-12

CD-11 Indicator

Press the G/N/T/

Menu F xx is displayed..... This is the full scale capacity

setting. xx= value last set. Available selections are are shown in the Setup
Table in section 2.6.3.

Press the PRINT/

Units button until desired capacity value is reached.

Press the G/N/T/

Menu button, gd0.001 is displayed. This is the graduation

size setting. Available selections are shown in the Setup Table in section
2.6.3.

Press the PRINT/

Units button until desired graduation value is reached.

Press the G/N/T/

Menu button, CP 5 kg is displayed. This is the full scale

calibration point setting. The range is from 5kg/lb to 100% Full scale capacity.

Press the PRINT/

Units button until desired calibration value is reached.

Press the G/N/T/

Menu button to end this block, SAVE is displayed.

Press the G/N/T/

Menu button to save the menu setup setting. The next menu

rEAD is displayed. The Indicator is now matched up with the scale base and
the Indicator parameters may now be set and calibrated.

or

Press the PRINT/

Units button to return to the SEtuP menu without saving

changes.

2.6.4 Setup Menu (Cont.)

Procedure