EN-15
CD-11 Indicator
Press the PRINT/
Units button until desired alternate mode setting is reached.
Alternate Mode enables either simple parts counting or display hold functions.
When ALtdH (display hold) is selected, a choice of manual, semi-automatic
or automatic settings are available. The alternate mode can be turned off so
that neither mode is available. It is not possible to have both modes activated
at the same time. For a complete description of alternate modes, refer to
Section 4 Operation.
2.6.5 Readout Menu (Cont.)
ALTERNATE MODE - not available with LFT ON or CAN
Press the G/N/T/
Menu button, ALtOFF is displayed. This is the alternate mode
setting.
Selections are:
OFF Standard weighing (this is the default setting)
PC Parts Counting
DH Display Hold - Man (manual)
Semi (semi-automatic)
Auto (automatic)
AUTO ZERO
Press the G/N/T/
Menu button, AZt 0.5 is displayed. This is the Auto Zero
Threshold setting. Selections are:
0.5d Sets threshold to 0.5 divisions. (this is the default setting)
1d Sets threshold to 1 division.
3d Sets threshold to 3 divisions.
Auto Zero minimizes the effects of temperature changes and small distur-
bances on the zero reading. The Indicator maintains the zero display until the
threshold is exceeded.
Press the PRINT/
Units button until desired auto zero threshold setting is
reached.
AUTO OFF TIMER
Press the G/N/T/
Menu button, AOtOFF is displayed. This is the Auto Off Timer
setting. When set ON, the Indicator will shut off automatically after 5 minutes
has elapsed on the condition that no button is pressed and the scale base
platform is stable during that period.
Press the PRINT/
Units button for selections ON or OFF. OFF is the default setting.