Ohaus CD-11 Indicator Manual multi User Manual
Page 22
EN-16
CD-11 Indicator
2.6.5 Readout Menu (Cont.)
RETAIN ZERO DATA
Press the G/N/T/
Menu button, Un rZdOFF is displayed. This is the Retain Zero
Data setting. When set On, the Indicator stores the current zero point and
restores it on the power-up.
Press the PRINT/
Units button for selections ON or OFF. OFF is the default
setting.
LCD BACK LIGHT
Press the G/N/T/
MENU button, bLAutO is displayed. This is the LCD backlight
setting. Selections are:
Auto Turns off the backlight in 5 seconds (this is the default setting)
ON Backlight is on continuously
OFF Backlight does not turn on
Press the PRINT/
Units button until desired LCD backlight setting is reached.
EP
This is a service function and is not a user-operated feature. OFF is the default
setting.
Not available with LFT ON or CAN.
SAVE
Press the G/N/T/
MENU button to end this block, SAVE is displayed.
Press the G/N/T/
MENU button to save the readout menu settings. The next
menu Print is displayed.
or
Press the PRINT/
Units button to go back to the SEtuP menu without saving
changes.
NOTE: If initial setup, go to the next paragraph. To exit from the SEtuP menu,
press the PRINT/
Units button to skip to PRINT then to LOCKSW, then Quit. Press
the G/N/T/
MENU button to go back to the weighing mode.