beautypg.com

Error codes list – Ohaus GT4100DG Electronic Balances User Manual

Page 46

background image

46

Error Codes List

Error Codes List

Error Codes List

Error Codes List

Error Codes List

The following list describes the various error codes and which can appear on the
display and the suggested remedy.

Data Errors

Data Errors

Data Errors

Data Errors

Data Errors

0.0

Transient error (hardware error, probably static discharge). If error persists,
the balance must be serviced.

Tare Errors

Tare Errors

Tare Errors

Tare Errors

Tare Errors

2.0

Balance is unable to stabilize within time limit after taring. Environment is too
hostile or balance needs recalibration.

2.1

Power on load out of specification.

Calibration Errors

Calibration Errors

Calibration Errors

Calibration Errors

Calibration Errors

3.0

Incorrect or no calibration weight used for calibration. Recalibrate with correct
weights.

RS232 Errors

RS232 Errors

RS232 Errors

RS232 Errors

RS232 Errors

4.0

Bad RS232 frame. Check RS232 menu parameters and correct.

4.4

RS232 buffer is full (if installed). May occur if no printer or computer is connected
to the interface. To clear buffer, turn balance off or enter Print menu and select
END.

4.5

Function is disabled by the Lock switch.

User Errors

User Errors

User Errors

User Errors

User Errors

7.1

Bad percent (%) mode, sample too low.

7.2

Number outside of display capacity.

Over-Under Load Errors

Over-Under Load Errors

Over-Under Load Errors

Over-Under Load Errors

Over-Under Load Errors

8.0

Hardware error causing an internal weight signal which is too low. Check if
platform or platform support is off. If not, the balance must be serviced.

8.1

Hardware error caused by an internal weight signal which is too high. Check
load on the platform which may be excessive. If error persists, the balance must
be serviced.