User menu – Ohaus GT4100DG Electronic Balances User Manual

Page 21

background image

21

Cal Test (Cont.)

Cal Test (Cont.)

Cal Test (Cont.)

Cal Test (Cont.)

Cal Test (Cont.)

4.

, no mass on the platform.

5.

, value of mass which

must be placed on the platform.

6.

allows the selection of either 25%, 50%, 75%, or 100% of full

span to be used to calibrate the balance.

5. Place required mass on platform.

6.

. While -C- is

displayed, the balance weighs the test mass based on current calibration data,

then displays the difference between the measured value and requested value.

The display shown is an example of when the test mass equals the mass value

stored in memory.

7. Remove mass from platform. After a short period of time, the balance returns

to the weighing mode.

USER MENU

USER MENU

USER MENU

USER MENU

USER MENU

RESET

YES

YES

YES

YES

YES/NO

AL

0, 1

1

1

1

1, 2, 3

STABILITY

.5d

5d

5d

5d

5d, 1d, 2d, 5d

AUTO ZERO

OFF, .5d

.5d

.5d

.5d

.5d, 1d,

3d

BEEP

ON/OFF

OFF

OFF

OFF

OFF

END, User

USER MENU

USER MENU

USER MENU

USER MENU

USER MENU

The User menu is used to adapt the balance to environmental conditions. It
contains submenus which enable you to turn features on or off, and program
balance parameters.

Reset

changes all submenus to original factory default

settings.

Reset

does not appear if menu has been locked out.

AL

specifies the

averaging level.

STB

specifies the desired stability range.

Auto Zero

sets the

automatic zero threshold.

Beep,

when set on, provides audible tones to signify

various balance conditions.

End User

is used to exit the Setup menu and store

the selections. The following figure illustrates the sequence in which submenus
appear on the User menu. Items shown in bold type are the default settings.