beautypg.com

A.2 auxiliary connector j8 pinout, Appendix – American Magnetics 187 Self-Compensating Liquid Level Controller User Manual

Page 69

background image

58

Appendix

Auxiliary connector J8 pinout

A.2 Auxiliary Connector J8 Pinout

The HI level and LO level contacts are provided for external use by the
customer. When a HI or LO level condition exists, the respective contact
pairs are closed. The HI and LO setpoints provide 0.05% hysteresis,
however the respective contact pairs still may “chatter” if the liquid
sloshes, bubbles, etc.

The HI level and LO level contacts also provide positive indication of a
power-off condition. With a power-off condition, both the HI level and LO
level contacts will be closed, which is a state unique to the power-off
condition.

The following table provides the specifications for the relay contacts:

Pin

Function

1

Not used

2

4-20 mA current loop output

(optional feature)

3

0-10 VDC output

(optional feature)

4

Analog output common

5 & 6

Lo level relay contacts (dry)

7 & 8

Hi level relay contacts (dry)

9

Not used

Max switching VA

10

Max switching voltage

200 VDC

Max switching current

0.5 A

Max continuous current

1.5 A

Dielectric between

contacts

200 VDC minimum

J8