Appendix
Auxiliary connector J8 pinout
A.2 Auxiliary Connector J8 Pinout
The HI level and LO level contacts are provided for external use by the
customer. When a HI or LO level condition exists, the respective contact
pairs are closed. The HI and LO setpoints provide 0.05% hysteresis,
however the respective contact pairs still may “chatter” if the liquid
sloshes, bubbles, etc.
The HI level and LO level contacts also provide positive indication of a
power-off condition. With a power-off condition, both the HI level and LO
level contacts will be closed, which is a state unique to the power-off
condition.
The following table provides the specifications for the relay contacts:
Pin
Function
1
Not used
2
4-20 mA current loop output
(optional feature)
3
0-10 VDC output
(optional feature)
4
Analog output common
5 & 6
Lo level relay contacts (dry)
7 & 8
Hi level relay contacts (dry)
9
Not used
Max switching VA
10
Max switching voltage
200 VDC
Max switching current
0.5 A
Max continuous current
1.5 A
Dielectric between
contacts
200 VDC minimum
J8