beautypg.com

2 selecting the appropriate units, 3 configuring the length settings, Selecting the appropriate units – American Magnetics 187 Self-Compensating Liquid Level Controller User Manual

Page 25: Configuring the length settings, Operation

background image

14

Operation

Operational Guidelines

3.1.2

Selecting the Appropriate Units

Repeatedly press the

81,76

key to cycle between display units of percent,

inches, or centimeters. The level reading displays the sensor length, or
percentage of the measured span, above the zero offset that is immersed in
liquid (see diagram below).

The selected units may be changed at any time except during the process
of entering a numerical value.

3.1.3

Configuring the Length Settings

The Model 187 must be configured with appropriate values for the zero
offset
and measured span settings. AMI records the overall length for the
sensor on the first page of this manual as well as on the calibration label
affixed to the bottom of the Model 187 chassis. AMI has also preset the
minimum zero offset and maximum measured span to match the
purchased sensor.

The zero offset and measured span settings can be modified by the user
and are accessible by pressing the

‡/(1

key.

Note

The LEN (length) menu adjustments can only be performed in the
inches or centimeter units modes. The length adjustments are
inactive if the units are set for %, and the instrument will beep
indicating that the LEN menu cannot be entered.

MAXIMUM

MEASURED

SPAN

MINIMUM

ZERO

OFFSET

OVERALL

SENSOR

LENGTH

Overall
Sensor

Length

This is the total active region of the concentric-
tube, capacitance-based probe. This value is
recorded on the first page of this manual and on
the instrument’s calibration label.

Zero

Offset

Measured from the bottom of the probe to a
minimum of just above the liquid reference plates.
The liquid reference plates must remain
submerged at all times for proper operation.

Measured

Span

Measured from the top of the zero offset (zero
point) to a maximum of just below the vapor
reference plates. The vapor reference plates must
remain exposed at all times for proper operation.
Levels are expressed as a percentage of the
measured span submerged in liquid.