Remote Interface Reference

Remote Units Commands and Queries

W

CALibration:SELect {0|1}

or

{USER|FACTORY}, <

cal_passcode

>

Selects between the USER and FACTORY calibrations. For example, the
command:

CAL:SELECT USER, 1234;

selects the USER calibration (if available) and requires that the
calibration passcode (“1234” in this example) be provided as a parameter.
If the calibration passcode is missing or incorrect, the Model 187 will beep
indicating an error. Also, the Model 187 will only allow selection of the
USER calibration if the user has previously performed a USER calibration
via the front panel interface or remotely.

W

CALibration?

Returns the currently selected calibration.

0,”USER”

is returned if the USER calibration is currently selected.

1,”FACTORY”

is returned if the FACTORY calibration is selected.

W

CALibration:PERForm <

cal_passcode

>

Performs a USER calibration if the correct calibration passcode is provided
as a parameter. The USER calibration is also made the current calibration
selection upon completion.

4.5.8

Remote Units Commands and Queries

The remote units commands and queries provide the capability of selecting
between units of percent, inches, and centimeters for all parameters and
return values for level and alarm commands and queries. The remote units
are independent of the display units selected via the front panel. The
remote units setting is persistent between power-ups of the Model 187.

W

UNITs {0|1|2}

Selects between the percent (0), inches (1), and centimeters (2) units. For
example, the command:

UNITS 1;

selects inches for the remote units.

W

UNITs?

Returns the currently selected units.

0,”PERCENT”

is returned if the remote units are percent.