Remote Interface Reference

Error Messages

4.6.1

Command Errors

-101,"Unrecognized command"

The command string sent was not identified as valid. Check the command
string for invalid characters or separators, syntax errors, or for errors in
the mnemonics. Spaces are not allowed before or after colon separators,
and at least one space must separate a command string from the
parameter(s).

-102,"Invalid argument"

The argument provided as a parameter for the command was invalid.
Value arguments must be of the following form:

W

an optional plus or minus sign,

W

a sequence of decimal digits, possibly containing a single decimal
point, and

W

an optional exponent part, consisting of the letter

e

or

E

, an optional

sign, and a sequence of decimal digits.

Enable_value arguments must be within the inclusive range of 0 to 255.

-103,"Non-boolean argument"

The command required a parameter in the form of 0 or 1. No other form of
the parameter is allowed.

-104,"Missing parameter"

The command required at least one argument which was not found before
the termination character(s).

-105,"Out of range"

At least one of the parameter values received was out of the valid range.
Refer to the Menu Reference section beginning on page 18 for the
acceptable ranges for each instrument setting.

-106,"Attempted to set length in percent"

The user attempted to set the zero offset or measured span in the remote
units of percent. The remote units must be inches or centimeters.

-107,"Incorrect passcode"

The user attempted to modify a calibration setting but failed to provide the
correct calibration passcode.

-108,"USER cal not available"

The user attempted to select the USER calibration when none exists. The
operator must perform a USER calibration in order to enable selection.